With a microwave disinfector
Microwave disinfectors ensure effective disinfection of the
product components in a standard commercial microwave
oven.
NOTICE
The LC interrupter is not designed to withstand exposure to
microwaves, and must therefore not be disinfected in a mi
crowave oven.
Place the product components in
the disinfector with enough space
between them to allow the steam
to reach all surfaces reliably.
Maintain the following times when disinfecting:
Wattage
Disinfection time
Cooling time
850 – 950
4 min.
2 min.
600 – 850
6 min.
2 min.
Information about the amount of water required and about the
disinfection process is included in the instructions for use of the
devices in question (disinfector and microwave).