background image

QUESTIONS ABOUT ERAPID?  CALL 1-866-335-6943  

|  

14

  

  

I

 

USING THE OPTIONAL EASYCARE

EASYCARE (Fig. 30), available separately, is a cleaning aid 
designed to maintain the standard performance of the eRapid 
Aerosol Head. It is important not to use EASYCARE more than 2 
times per week to maintain maximum eRapid Aerosol Head life. 
1.  Read the Instructions For Use for EASYCARE carefully and 

completely before using it for the first time.

2.  Connect EASYCARE to the Controller as described in the 

EASYCARE Instructions For Use.

3.  To start the EASYCARE cleaning cycle press and hold down  

the On/Off button on the Controller for at least 10 seconds. 
First you will hear one “beep” and the status light will 
illuminate green. The START screen (Figure 17a) will appear  
on the display.

4.  After about 10 seconds you will hear a second “beep” and the 

screen will change showing that the EASYCARE cleaning has 
started (Figure 31). Release the On/Off button.

NOTE: During EASYCARE cleaning make sure that either one of 

the EASYCARE screens (Figure 31) are showing in the Controller 

display. If the TREATMENT screens show (Figure 17b) shut 

the Controller off and restart the EASYCARE cleaning cycle in 

accordance with the above instructions.

5.  When the EASYCARE cleaning cycle is completed you will hear 

two “beeps”, and the same DOSE COMPLETE screen will appear 
(Figure 18). The Controller will automatically shut off.

NOTE: In the event of a malfunction, see the Message Screens 

and Warnings chart on Page 11, locate the alternating symbols, 

and take the corrective action noted.

6.  Rinse (see Section H) and Disinfect (see Section J) the 

Aerosol Head after every EASYCARE cleaning cycle.

7.  Refer to the EASYCARE Instructions For Use in disassembling 

and maintaining the EASYCARE.

Figure

30

Figure

32

Figure

31

With Battery

With AC Adapter

J

 

EFLOW NEBULIZER HANDSET DISINFECTION

WARNING 

To reduce the risk of serious or fatal illness caused by 
contamination of the eRapid Nebulizer Handset you must 
also 

DISINFECT THE HANDSET AT THE END OF EVERY 

TREATMENT DAY.

 In addition, the eRapid Nebulizer Handset is 

strictly FOR SINGLE PATIENT USE ONLY.

NOTE: The disinfection instructions described may be used in 

both home and institutional environments.

1.  Clean the eRapid Nebulizer Handset as described in Section H.
2.  AT THE END OF EVERY TREATMENT DAY, disinfect the 

eRapid Nebulizer Handset with a commercially available 
medical device disinfectant such as Control III®. 

Follow the 

disinfectant manufacturer’s directions for use.

3.  The eRapid Nebulizer Handset may also be disinfected by 

boiling the parts, including the Aerosol Head, in a clean pot of 
DISTILLED water for a full 5 minutes. 

Do not boil in regular tap 

water, as it may damage the Aerosol Head.

4.  Air dry the eRapid Nebulizer Handset parts on a clean, lint-free 

towel (Figure 32).

5.  After the eRapid Nebulizer Handset parts are completely 

dry, place them in a bag for storage. DO NOT re-assemble the 
eRapid Nebulizer Handset until just prior to the next treatment. 
This is  to ensure that the silicone seals do not become 
deformed  over time.

6.  Hospitals and clinics can sterilize the eRapid Nebulizer 

Handset parts by wet autoclaving at 250° F (121° C) for 30 
minutes. 

Follow the autoclave manufacturer’s instructions 

 for  use.

Содержание 178 G 1030 eRapid

Страница 1: ...WARNING ReadandunderstandtheseInstructions forUseandallsafetyprecautionsit contains Improperusecancauseserious orfatalinjuryand orillness CAUTION USFederallawrestrictsthis devicetosalebyoronorderofap...

Страница 2: ...3 2 www pari com register 1 866 335 6943 Registration card enclosed REGISTER FIRST WARRANTY ACTIVATION REQUIRED Three different options to choose from Conforms to UL 60601 1 Certified to CAN CSA C22 2...

Страница 3: ...precautions for the eRapid Nebulizer System Save these Instructions For Use for future reference Before using your eRapid Nebulizer System you should read and understand the entire Instructions For U...

Страница 4: ...ratewhereoxygenisbeingadministeredin aclosedenvironmentsuchasanoxygentent QUESTIONS ABOUT ERAPID CALL 1 866 335 6943 4 B SAFETYPRECAUTIONSANDIMPORTANTSAFEGUARDS 4 Alwaysunplugthisproductimmediatelyaft...

Страница 5: ...ENDOFEVERYTREATMENTDAY See SectionJfordisinfectioninstructions 14 Thisproductcontainssmallpartsthatmaypresenta chokinghazardtosmallchildren TheeRapidNebulizer ConnectionCordalsopresentsastrangulationh...

Страница 6: ...g items Note that two complete Nebulizer Handsets are included FOUR AA BATTERIES NEBULIZER CONNECTION CORD EBASE CONTROLLER AEROSOL CHAMBER MOUTHPIECE WITH EXHALATION VALVE AC POWER SUPPLY AEROSOL HEA...

Страница 7: ...ife of 90 minutes of use is based on disposable batteries that meet the specifications listed in Specifications on the back cover of these Instructions For Use E USINGACPOWERSUPPLY 1 The Controller ca...

Страница 8: ...RIO Investigational eFlow Being careful to touch ONLY the plastic outer ring of the Aerosol Head Figure 7 press the two flexible plastic tabs on the side of the Aerosol Head while firmly inserting it...

Страница 9: ...9 QUESTIONS ABOUT ERAPID CALL 1 866 335 6943 Figure 7 Figure 9 Figure 10 Figure 11 Figure 12 Figure 8 Figure 7a Figure 7b AEROSOL HEAD DETAIL TABS EXHALATION VALVE a b...

Страница 10: ...he treatment WARNING KeeptheeRapidNebulizerHandsetlevelwheninuse Ifthe handsetistilted thedosagemaynotbeaccurate 8 Breath normally in and out through the Mouthpiece Continue to inhale and exhale comfo...

Страница 11: ...sound When the alternating screens appear follow the text in the Description Remedy to the right of the screen symbols below SYMBOL ALTERNATING SCREENS DESCRIPTION REMEDY BATTERY LEVEL LOW Replace wi...

Страница 12: ...n additives harmful to the eRapid Nebulizer Handset Also Do not put the Aerosol Head in a microwave oven Do not attempt to clean the Aerosol Head in a dishwasher Mechanical cleaning of the Aerosol Hea...

Страница 13: ...13 QUESTIONS ABOUT ERAPID CALL 1 866 335 6943 Figure 22 Figure 23 Figure 26 Figure 27 Figure 28 Figure 29 Figure 24 Figure 25 TAB a b a b...

Страница 14: ...ols and take the corrective action noted 6 Rinse see Section H and Disinfect see Section J the Aerosol Head after every EASYCARE cleaning cycle 7 Refer to the EASYCARE Instructions For Use in disassem...

Страница 15: ...ck to see if the alternating BATTERY LEVEL LOW icons are flashing on the controller display If they are replace the batteries or use the AC Power Supply Answer If you are using the AC Power Supply che...

Страница 16: ...arge Time up to 15 hours depends on type of battery charger DISPOSABLE BATTERIES Panasonic Evoia or equivalent 1 5V Panasonic Power Digital or equivalent 1 5V MANUFACTURED IN GERMANY DISTRIBUTED BY PA...

Отзывы: