FILTER/VALVE SET

6 MONTH

WARRANTY!

PART # 041F05

INSTRUCTIONS FOR USE

FOR SINGLE PATIENT USE (ONE PERSON) ONLY

INSPIRATORY VALVE CAP INCLUDED WITH LC PLUS® NEBULIZER

FILTER

Y PIECE

STANDARD 
MOUTHPIECE

II. CLEANING THE 

PARI

 FILTER/VALVE SET

WARNING

To reduce the risk of infection, illness, or injury from contamination, you must thoroughly clean 
all parts of the nebulizer as instructed in the nebulizer IFU.

1.  Remove the PARI filter/valve set from the nebulizer and disassemble all parts.  Open the filter 

and remove filter pad. Discard filter pad. 

2.  Wash all parts of the PARI filter/valve set with warm water and liquid dish soap. 
3.  Rinse thoroughly with warm water and shake out water. 
4.  Air dry or hand dry PARI filter/valve set parts on a clean, lint free cloth. 
5.  You can also wash all parts of the PARI filter/valve set in a dishwasher.  Place the PARI  

filter/valve set parts in a dishwasher basket, then place on the top rack of the dishwasher.  
Remove and dry the parts when the cycle is complete. 

III. DISINFECTING AND STERILIZING THE 

PARI

 FILTER/VALVE SET

WARNING: TO REDUCE RISK OF INFECTION:

1.  PARI filter/valve set for single patient use (one person) only.  Do not share with  

other patients. 

2.  You must regularly disinfect or sterilize the PARI filter/valve set between treatments.  

Failure to do so could lead to serious or fatal illness. 

For patients with Cystic Fibrosis, the CF Foundation recommends boiling for a full 5  
minutes to disinfect.  

1.  Clean the PARI filter/valve set as described in section II. Every other treatment day, soak all parts 

of the PARI filter/valve set (except filter pad) in a solution of 1 part distilled white vinegar and 
3 parts hot tap water for 1 hour. You can substitute respiratory equipment disinfectants (like 
Control III®) for distilled white vinegar (follow manufacturer’s instructions for mixing). Rinse all 
parts of the PARI filter/valve set thoroughly with warm tap water and dry with a clean, lint-free 
cloth.  Discard the vinegar solution when disinfection is complete.  

2.  The PARI filter/valve set parts may also be disinfected by boiling them in water for a full 5 

minutes. Hospitals and clinics can sterilize the PARI filter/valve set parts by autoclaving (a 
maximum of 277°F), follow autoclave manufacturer’s instructions. Dry parts on a clean,  
lint-free cloth. 