PARI
FILTER/VALVE SET
6 MONTH
WARRANTY!
PART # 041F05
INSTRUCTIONS FOR USE
FOR SINGLE PATIENT USE (ONE PERSON) ONLY
INSPIRATORY VALVE CAP INCLUDED WITH LC PLUS® NEBULIZER
FILTER
Y PIECE
STANDARD
MOUTHPIECE
II. CLEANING THE
PARI
FILTER/VALVE SET
WARNING
To reduce the risk of infection, illness, or injury from contamination, you must thoroughly clean
all parts of the nebulizer as instructed in the nebulizer IFU.
1. Remove the PARI filter/valve set from the nebulizer and disassemble all parts. Open the filter
and remove filter pad. Discard filter pad.
2. Wash all parts of the PARI filter/valve set with warm water and liquid dish soap.
3. Rinse thoroughly with warm water and shake out water.
4. Air dry or hand dry PARI filter/valve set parts on a clean, lint free cloth.
5. You can also wash all parts of the PARI filter/valve set in a dishwasher. Place the PARI
filter/valve set parts in a dishwasher basket, then place on the top rack of the dishwasher.
Remove and dry the parts when the cycle is complete.
III. DISINFECTING AND STERILIZING THE
PARI
FILTER/VALVE SET
WARNING: TO REDUCE RISK OF INFECTION:
1. PARI filter/valve set for single patient use (one person) only. Do not share with
other patients.
2. You must regularly disinfect or sterilize the PARI filter/valve set between treatments.
Failure to do so could lead to serious or fatal illness.
For patients with Cystic Fibrosis, the CF Foundation recommends boiling for a full 5
minutes to disinfect.
1. Clean the PARI filter/valve set as described in section II. Every other treatment day, soak all parts
of the PARI filter/valve set (except filter pad) in a solution of 1 part distilled white vinegar and
3 parts hot tap water for 1 hour. You can substitute respiratory equipment disinfectants (like
Control III®) for distilled white vinegar (follow manufacturer’s instructions for mixing). Rinse all
parts of the PARI filter/valve set thoroughly with warm tap water and dry with a clean, lint-free
cloth. Discard the vinegar solution when disinfection is complete.
2. The PARI filter/valve set parts may also be disinfected by boiling them in water for a full 5
minutes. Hospitals and clinics can sterilize the PARI filter/valve set parts by autoclaving (a
maximum of 277°F), follow autoclave manufacturer’s instructions. Dry parts on a clean,
lint-free cloth.
CONTROL III® IS A REGISTERED TRADEMARK OF MARIL PRODUCTS, INC
© 2016 PARI Respiratory Equipment, Inc. DS-41-05 REV F 4-16