Gener
al
Information
Pr
epar
e
Oper
ate
Help
Technology
Electromagnetic compatibility (EMC)
146
21.2
Compliance level
The interference immunity levels in accordance with IEC 60601 are fulfilled.
Manufacturer's declaration – electromagnetic interference
The PR 40 is designed to be operated under the electromagnetic conditions specified below. The customer
or the user of the device should ensure that it is used under these conditions.
Emission measurements
Compliance
Electromagnetic conditions – guideli-
nes
HF emissions in accordance with CISPR11 Group 1
The PR 40 only uses HF energy for its
internal functions. Therefore its HF
emissions are very low, and it is unlikely
that adjacent electronic devices will be
disturbed by these.
HF emissions in accordance with CISPR11 Class B
The device is designed to be used in all
set-ups including living areas and those
which are directly connected to a public
power grid that also supplies power to
buildings being used for living purposes.
Emission of harmonics in accordance
with IEC61000-3-2
Class A
Emission of voltage fluctuations/flickers
in accordance with IEC61000-3-3
fulfilled
Table 201 in accordance with DIN EN 60601-1-2, 6.8.3.201