(c) Finding your Intensity level 

After setup, repeat for each treatment session 

 
1. Ensure electrode has a firm connection to the left tragus (See p.7 image). 
 
2. Using the up arrow on the left of the keypad, for channel 1, slowly 
increase the intensity by 1 level every 2 seconds to the point where a 
tingling is felt on the tragus, going to a maximum of Level 36.  
 
Take care in doing this as it can be difficult to perceive the stimulation at first 
and perception of sensation may be sudden. If the stimulation is painful at 
first perception, turn the device off, then turn the device on and repeat this 
step to find a lower intensity.  
 

Note: 

If no stimulation is felt at Level 36, 

apply a small amount of moisture to the 

tragus

 and repeat step 2 again making sure to start from Level 0. Altering the Hz 

(pulse rate), between 1 and 25 Hz can help with finding the sensory threshold, as 
well as moving the earclip into a more firm position. If all of the above fails, please 
contact us for further instructions. 

 

 
6. Once the sensory threshold has been found, decrease the intensity to 1 
level below this level. This is now set at the therapeutic level you can use for 
treatment. If sensitivity increases during the course of treatment, you can 
reduce intensity accordingly. 
 

 
 
 
 
 
 
 
 
 
 
 

12 

Содержание tVNS

Страница 1: ...Parasym tVNS Device Quick Start Guide...

Страница 2: ...1...

Страница 3: ...for you that you can use every day Two AA alkaline batteries are included so you can start straight away Before using your device please make sure you have read all the precautions warnings and safet...

Страница 4: ...the electrodes and accessories recommended by the manufacturer as using others may be unsafe Any electrodes that have current densities exceeding 2 mA cm 2 may require special attention of the operato...

Страница 5: ...or cracked The device casing is cracked dented or appears to be damaged Discontinue use if you experience Lightheadedness dizziness or chest pain Excessive skin irritation Caring for Your Device Do no...

Страница 6: ...6 of participants discontinued use due to side effects where the most common side effect reported was skin irritation at the stimulation site which typically ceased shortly after use The following si...

Страница 7: ...Instructions For Use 6...

Страница 8: ...7...

Страница 9: ...Electrode Placement Connect the electrode to the tragus of the left ear running the cable behind the neck over the right shoulder 8...

Страница 10: ...9...

Страница 11: ...into channel 1 3 Electrode Placement 3 1 Clip the electrode to your left tragus See diagram on Page 8 Make sure this is a good connection with both gold plates having a firm contact with the skin 4 Pu...

Страница 12: ...s of use per day This can be achieved by pressing the timer button to 60 min The device will automatically shut down after 60 minutes Please find a convenient time each day using the treatment at the...

Страница 13: ...ception turn the device off then turn the device on and repeat this step to find a lower intensity Note If no stimulation is felt at Level 36 apply a small amount of moisture to the tragus and repeat...

Страница 14: ...r treatment session should be one level below the sensory level where it can be perceived or felt For most people this is between Level 1 Level 15 however some people need to go to the maximum to find...

Страница 15: ...ectedly This is likely to be the power save function If no buttons are pressed and the intensity level is zero the unit will power off automatically after a few minutes There is no stimulus simulation...

Страница 16: ...Visit our Help Center Our Help Centre has answers to most common questions You can access this at www parasym co hc 15...

Страница 17: ...ort team Reach out to us via our support messaging widget on our website www parasym co and we ll get right back to you You can also email us directly at hello parasym co and this will also go through...

Страница 18: ...mmerse the device in water 4 Do not use any other cleaning solution than soap and water Testing the Device The device can be tested by making sure the batteries are in the device correctly pressing th...

Страница 19: ...s shown in the image on page 8 Note the safety and efficacy of the Parasym device has not been established in individuals with Implantable medical devices a history of cardiac disease or arrhythmia di...

Страница 20: ...ctangular Mode 5 Modes are available Mode A Constant Stimulation Mode B Burst I Two train per second 250ms on 250ms off 25 pulses within one train pulse width 200 S Mode C Burst II Two train per secon...

Страница 21: ...ects on the environment Symbols Explanation Type BF Applied Part Refer to Instructions for Use Compliance with WEEE regulations Manufacturer European Representative IP22 Protected against access to ha...

Страница 22: ...Version 4 2 www parasym co 21...

Отзывы: