REGULAR INSPECTION AND DAILY MAINTENANCE
1. REGULAR INSPECTION
▪
To ensure safe use, inspect the product (at least once a year). The product exhibits wear and
deterioration according to the frequency and conditions of use. Inspect for loose parts, the
operation of casters, and the presence of damage regularly.
▪
If you notice an irregularity or want a detailed inspection, contact your dealer or Paramount Bed
dealer.
2. DAILY MAINTENANCE
▪
To remove dirt from parts
Complete steps (1) to (3) listed below.
(1) Wipe off parts using a soft cloth soaked in a neutral cleaning solution lightly diluted with water that has
been wrung out to remove excess fluid.
(2) Ring out as much moisture from the cloth as possible and then wipe up any remaining cleaning solution
residue.
(3) Use a dry cloth to thoroughly wipe away any leftover moisture.
▪
Disinfecting parts
Use one of the cleaning solutions listed below at the specified concentrations, making sure to follow the
instructions in the manual or other materials available for the cleaning solution.
76.9% - 81.4% ethanol for disinfection
0.05% - 0.2% benzalkonium chloride (e.g. Osvan)
0.05% - 0.2% benzethonium chloride (e.g. Hiamine)
0.05% chlorhexidine gluconate (e.g. Hibitane)
0.02% - 0.05% sodium hypochlorite (e.g. Milton)
CAUTION
●
Do not use volatile cleaning fluids (such
as paint thinner, benzene, or gasoline)
and cresol. Discoloration or deterioration
could result.
●
Do not use cleaning solutions other than
those listed above with the product.
Damage or deterioration could result.
●
Thoroughly dry off parts after using a neutral cleaning solution. Failure to
do so could cause resin parts to crack.
●
Do not use a fluorescent brightener. A change in gloss or discoloration could result.
The
●
Product cannot be used with an ozone sterilizer, an autoclave, etc.