Insert the oxygen cylinder holder into any of the
optional accessory attachment holes on the two
corners of head end of the medical device.
Do not attach the oxygen
cylinder holder to the foot end
of medical device. When the
medical device is operated
with reverse trendelenburg,
there is a risk that the foot may
be caught between the oxygen
cylinder holder and the floor.
2.
Align a groove of the oxygen cylinder holder with
the bolt in the bottom of the optional accessory
attachment hole.
●
Use the oxygen cylinder
holder at position A in the
figure on the left. If used in
position B or C, the medical
device may become
unstable, fall over, and
cause injury.
●
Do not install the IV pole on
the right side (left side) of
the medical device when
using the oxygen cylinder
holder on the right side (left
side) of the medical device.
If used in one side of the
medical device, the medical
device may become
unstable, fall over, and
cause injury.