PARAGONCARE ACERO Скачать руководство пользователя страница 18

USER GUIDE  |  ACERO ELECTRIC COUCH

14

9. Dimensions & Data

Dimensions

Overall Width

700mm

Overall Width (with rails)

750mm

Overall Length 3 section

1940mm

Overall Length 2 section

1965mm

Height to Platform Top

480mm – 880mm

Backrest Adjustment – 2 section

-24 degrees to +84 degrees

Backrest Adjustment – 3 section

0 degrees to 80 degrees

Footrest Adjustment – 3sec only

-24 degrees to +84 degrees

Safe Working Load

200Kg

Product weight (approx)
Without accessories

72kg

Electrical Data (230V)

Power in

2A max. at 230V~ , 50Hz.

Duty rating

Intermittent 10%, Max. 2min. / 18 min.

Electrical Safety Standard

Complies with EN 60601-1/ UL60601-1

Electric shock protection

Class 2  

EMC

Complies with BS EN 60601-1-2 2001

Liquid ingress protection

IP54

Internally Powered Equipment (Emergency Battery backup only)

Source

2 x 12V series connected Batteries

Type

Sealed, rechargeable lead/acid gel 1.2 amp hrs.

Содержание ACERO

Страница 1: ...ELECTRIC EXAMINATION COUCH User Guide Paragon Care Group Pty Ltd ABN 44 136 627 971 11 Dalmore Drive Scoresby VIC 3179 Australia T 1300 369 559 F 03 8833 7890 info paragoncare com au paragoncare com...

Страница 2: ......

Страница 3: ...Contents Introduction 2 Clinical Applications 3 Installation 4 Operation 5 Cleaning 8 Maintenance 9 Troubleshooting 12 Warranty and Service 13 Dimensions Data 14 Transportation Storage 15...

Страница 4: ......

Страница 5: ...lease contact 1300 369 559 to arrange an appointment Warnings given in this manual identify possible hazards in procedures or conditions which if not correctly followed could result in death injury or...

Страница 6: ...es dependent upon couch configuration are available within the range Fully Electric Available as 2 section and 3 section Electric Height adjustment Electric Backrest and Footrest adjustment Hand held...

Страница 7: ...physio centers Doctors surgeries Indications The use of this examination treatment couch is to assist in the patient positioning and treatment whilst reducing lifting and handling stresses for the tr...

Страница 8: ...uipment Wrap the electrical cable around the cable retainer only when couch is to be moved Cable retainer is located on side of base frame Always hang handset on handset holder located on both sides o...

Страница 9: ...brakes when the couch is stationary Before operating the couch make sure the patient is positioned correctly to avoid entrapment or imbalance Lower the couch to the minimum height when the patient is...

Страница 10: ...vidual locking brake castors To operate braking push down on the brake activation lever on each castor To disengage the brake push down on the release lever situated above the brake activation lever C...

Страница 11: ...ation To raise the side Hold the top rail towards the middle swing outwards and raise to a vertical position and make sure that both pivot posts drop back inside pivot mount slots 4 CAUTION Do NOT use...

Страница 12: ...ospital s recommended cleaning solution Take extra care with areas that may harbour dirt or dust Rinse with clean water and dry with paper towels To remove potentially infectious material such as body...

Страница 13: ...f and dirt from the axle Check tyres for wear and that the braking mechanism is functioning properly Base top pivot bolts shafts Check for excessive wear and tightness of fittings Replace if necessary...

Страница 14: ...supply for an uninterrupted 24 hour period 2 Disconnect the couch from the mains supply 3 Apply a load of 80kg to platform 4 Raise and lower the platform from minimum to maximum height three times 5 I...

Страница 15: ...ctors or insulation causing conductors to short together which could cause unwanted movement of the bed Any handsets are found with obvious damage to the cable insulation immediately report the damage...

Страница 16: ...Check the handset plug is fully inserted in the control box Blown fuse Contact PargonCare Service Department on 1300 369 559 Damaged handset cable Replace handset One actuator fails to work Actuator...

Страница 17: ...request These contain full details of warranty terms and do not limit the statutory rights of the consumer For service maintenance and any questions regarding this or any other ParagonCare product ple...

Страница 18: ...t Adjustment 3sec only 24 degrees to 84 degrees Safe Working Load 200Kg Product weight approx Without accessories 72kg Electrical Data 230V Power in 2A max at 230V 50Hz Duty rating Intermittent 10 Max...

Страница 19: ...Do not drop Avoid shock or violent impact The equipment should be stored in a clean dry and well ventilated area The following limits apply during transport and for a storage period up to 15 weeks Th...

Страница 20: ...Paragon Care Group Pty Ltd 11 Dalmore Drive Scoresby 3173 Victoria Australia T 1300 369 559 F 03 8833 7890 E info paragoncare com au paragoncare com au...

Отзывы: