![PAR PHYSIO-PORT Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/par/physio-port/physio-port_operators-manual_3549989024.webp)
Cleaning, Maintenance, Disposal
24
PHYSIO-PORT / PHYSIO-PORT AS
A8136 Revision A
6.4 Maintenance
Checks before each use
Before each use, visually check the device and the
cables for signs of mechanical damage.
If you detect damage or impaired functions which may
result in a hazard to the patient, the operator or third
persons, the device must be repaired before it can be
used again.
The PHYSIO-PORT devices represent a measuring
device, which is why they must be subjected to a
regular “Technical Safety Inspection" and “Technical
Inspections of the Measuring System" in accordance
with MPBetreibV § 11/ Annex 2. The time of the next
due inspection is indicated by the calibration seal (see
Fig. 2-1).
Technical Safety Inspections
For safety, the device requires regular maintenance.
To ensure functional and operational safety of the
PHYSIO-PORT devices, Technical Safety
Inspections should be carried out at least every two
years.
The checks are carried out within the framework of a
service agreement; please contact PAR Medizintechnik
Service for details.
The nature and extent of the inspections are described
in the corresponding sections of the Service Manual.
Upon request, PAR Medizintechnik provides a Service
Manual.
The device does not require any other maintenance.
Technical Inspections of the Measuring System
The non-invasive pressure measurement system of
the PHYSIO-PORT devices should be inspected
every two years.
The checks are carried out within the framework of a
service agreement; please contact PAR
Medizintechnik Service for details.
The nature and extent of the inspections are described
in the corresponding sections of the Service Manual.
Upon request, PAR Medizintechnik provides a Service
Manual.
6.5 Disposal of the Product
The products described in this operator
manual must not be disposed as unsorted
municipal waste and must be collected
separately. Please contact an authorized
representative of the manufacturer for
information concerning the
decommissioning of your equipment.
Caution
These checks shall be carried out by PAR
Medizintechnik or authorized companies.
Caution
These checks shall be carried out by PAR
Medizintechnik or authorized companies.