PHYSIO-PORT
User Manual Version 2017-11-12
- 31 -
Upward measurement on:
The inflation measurement method is activated (
Inflation
Measurement Technology, IMT)
Activate buzzer:
The internal buzzer(optional) can be switched on and off. If the
buzzer is on, than an acoustic signal sounds when the device is
switched on, before the measurement or in case of faulty
measurement.
Standard:
The
[Standard]
button will reset the parameters to the following values:
2 measurement Interval
Interval 1
(day phase):
06:00 - 21:59
exact every 15 min
Intervall 2
(night phase):
22:00 - 05:59
exact every 30 min
After saving the selected configuration, the
starting sequence of the recorder will
commence by clicking on the
[Start]
button.
The recorder will first check whether there
are any previous measurements that have
not yet been read. If this is the case, then a
warning will be displayed.
If this is not the case, then the recorder will
be switched off and the new patient data and
measurement intervals are stored in the
recorder.
CAUTION
•
Patient interchange
•
The recorder must be restarted on each new
measurement by Physioportwin software.
Otherwise, new measurements are assigned to the old
patient.
Only when a new long-term measurement is started,
will the recorder delete all old patient data and
measurement values.