Rotarod
16
ACCELERATION:
Figure 11. Screen in acceleration mode running
This is the second state after pressing the Hold Ramp button in this mode.
The motor ramps from 4 to 40 rpm, following the predetermined SLOPE TIME
once the Hold Ramp button has been pressed.
After the ramp is initiated, the ACCELERATION TIME cannot be changed.
The TIMER counters corresponding to the raised LEVERS, begin recording the
time after the Hold Ramp button is pressed.
Once the relevant LEVER is depressed, the TIMER will show the total rotating
time as the end point for that subject on that particular trial.
When the last subject falls from the rod, the instrument automatically enters
the STOP state.
Содержание 76-0770
Страница 31: ...Rotarod 30 04 06 2014...