![Panasonic MDF-U5586SC Скачать руководство пользователя страница 35](http://html.mh-extra.com/html/panasonic/mdf-u5586sc/mdf-u5586sc_operating-instructions-manual_267968035.webp)
34
SPECIFICATIONS
Product name
Ultra-Low Temperature Freezer
MDF-U5586SC
Ultra-Low Temperature Freezer
MDF-U7586SC
External dimensions
W890 mm x D870 mm x H1990 mm
W1130 mm x D870 mm x H1990 mm
Internal dimensions
W630 mm x D600 mm x H1280 mm
W870 mm x D600 mm x H1280 mm
Effective capacity
483 L
668 L
Exterior Painted
steel
Interior Painted
steel
Outer door
Painted steel
Inner door
ABS resin panel with stainless frame, 2 doors
Shelf
Stainless steel, 3 shelves (adjustable)
W608 mm x D535 mm, Load; 50
kg
/shelf
Stainless steel, 3 shelves (adjustable)
W848 mm x D533 mm, Load; 50
kg
/shelf
Access port
17 mm diameter, 2 locations (back, bottom left corner)
Insulation
Rigid polyurethane foamed-in place
Compressor
Hermetic type, Output; 1100 W (high stage side), 1100 W (low stage side)
Evaporator
Tube on sheet type
Condenser
Fin and tube type (high stage side), Shell and tube type (low stage side)
Refrigerant
R-404A (high stage side), R-508B (low stage side)
Temperature controller
Microcomputer control system
Temperature display
Digital display
Thermal sensor
Platinum resistance (Pt 1000
Ω
)
Alarm
High temp. alarm, Low temp. alarm, Power failure alarm
Filter check, Battery check, Fan motor check
Remote alarm contact
Allowable contact capacity: DC 30 V, 2 A
Battery
Nickel-metal-hydride battery, DC 6 V, 1100 mAh, Auto-recharge
Accessories
1 set of key, 1 scraper, 1 s
tick for air intake port cleaning
Weight
320
kg
365
kg
Voltage booster
Built-in
Optional component
Inventory rack (IR-220U)
Temperature re Recorder fixing (MTR-G85C*, MTR-85H+MDF-S3085)
Backup cooling kit (CVK-UB2/CVK-UB2I): LCO
2,
(CVK-UBN2): LN
2
Interface board (MTR-480, MTR-L03)
Data acquisition system (MTR-5000)
Note
:
• Design or specifications will be subject to change without notice.
• Refer to the updated catalog when ordering an optional component.
*: Power source of the temperature recorder shall be 220 V.
• When a data transmitting function to the PC is done, an interface board MTR-480 (option) and
communication cable of 9 pin Dsub cross type for RS232C are necessary.
• The connecting cable between the PC and MTR-L03 terminal is not enclosed. Provide a cross type
cable of less than 100 m length having a pin plug.