36
ALARM PARAMETER
1.
Press Menu key to lead the Menu screen.
2.
Press Tools #1 key to lead the Tools #1 screen.
3.
On the Tools #1 screen,
・
Press Alarm Setting #1 key to lead the Alarm
Setting #1 screen, it is possible to set automatic set
temperature alarm and automatic set CO
2
density
alarm (go to procedure
4
).
・
Press Alarm Setting #2 key to lead the Alarm
Setting #2 screen, it is possible to set alarm delay,
door alarm delay, ring back and remote alarm (go
to procedure
5
).
4.
On the Alarm Setting #1 screen, input each
parameter. Press Apply key to save the input value.
The display returns to the Tools #1 screen.
●
Each parameter setting
・
Temp. Alarm:
When the chamber temperature exceeds the scope,
the set temperature ± the set value of Automatic set
temperature alarm, the alarm is activated. Settable
range: ±1.0
o
C
~
±5.0
o
C, factory setting: ±1.0
o
C
。
・
CO
2
Alarm:
When the chamber CO
2
density exceeds the scope, the set CO
2
density ± the set value of Automatic set
CO
2
density alarm, the alarm is activated. Settable range: ±0.5 %
~
±5.0 %, factory setting: ±1.0 %
。