7.2 Advanced Settings (
)
7.2.1 Hold Settings (
)
This function sets the method for holding a judgment value (JUDGE.V).
To output normal measured values in real time, set the measurement mode to sample hold "S-
H" and leave external inputs turned OFF.
■
Measurement mode function (
)
This function allows you to select a hold mode from the following six types.
Measurement mode
(MEAS)
Function
Reference page
Sample hold
(S-H)
This function holds and outputs the judgment value (JUDGE.V)
only when external input is ON.
When external input is OFF, the judgment value (JUDGE.V) is
updated as necessary before being output. (Real-time output)
Peak hold (P-H)
When external input is turned ON, the maximum value during the
sampling period is held and output as the judgment value
(JUDGE.V). When external input is turned OFF, the maximum
value during the sampling period is continuously updated.
Bottom hold (B-H)
When external input is turned ON, the minimum value during the
sampling period is held and output as the judgment value
(JUDGE.V). When external input is turned OFF, the minimum value
during the sampling period is continuously updated.
Peak-to-peak hold
(P-P)
The difference between the maximum value and minimum value is
held and output as the judgment value (JUDGE.V).
NG hold (NG-H)
This function holds LOW judgment and HIGH judgment output.
Tab cancellation
(TB.CNCL)
This function holds and outputs the judgment value when the
amount of change in the judgment value exceeds the optionally set
“tab judgment threshold”.
The hold is released when the separately set “hold releasing
condition” is satisfied.
■
External trigger input function
Hold
If external trigger input turns ON, this function holds the judgment value (JUDGE.V) while the
input is ON.
7.2 Advanced Settings (
)
WUME-HGT-6
7-9
Содержание HG-T Series
Страница 2: ... MEMO 2 WUME HGT 6 ...
Страница 28: ... MEMO 2 8 WUME HGT 6 ...
Страница 96: ... MEMO 5 50 WUME HGT 6 ...
Страница 99: ... Note 2 Displayed only during calculation 6 1 Flow of Measurement Data WUME HGT 6 6 3 ...
Страница 128: ... MEMO 6 32 WUME HGT 6 ...
Страница 158: ... No timer OFF 7 2 Advanced Settings 7 30 WUME HGT 6 ...
Страница 177: ...Example Positioning a glass plate Setting method 7 2 Advanced Settings WUME HGT 6 7 49 ...
Страница 218: ...8 2 3 Side View Attachment HG TSV10 Units mm 8 2 Dimension Drawings 8 14 WUME HGT 6 ...
Страница 221: ...10 Troubleshooting 10 1 Troubleshooting 10 2 10 2 Error Messages 10 5 WUME HGT 6 10 1 ...
Страница 228: ...Menu Structure General Function Display Menu Structure General Function Display App 2 WUME HGT 6 ...
Страница 229: ...Menu Structure General Function Display WUME HGT 6 App 3 ...
Страница 230: ...Menu Structure Extended Function Display Menu Structure Extended Function Display App 4 WUME HGT 6 ...
Страница 231: ...Menu Structure Extended Function Display WUME HGT 6 App 5 ...
Страница 232: ... MEMO App 6 WUME HGT 6 ...
Страница 235: ... MEMO WUME HGT 6 ...
Страница 236: ...Panasonic Industrial Devices SUNX Co Ltd 2020 PRINTED IN JAPAN WUME HGT 6 ...