6
E
Table 2
Guidance and manufacturer’s declaration-electromagnetic immunity
CF-VCBC11U is intended for use in the electromagnetic environment specified below. The
customer or user of the CF-VCBC11U should assure that is used in such an environment.
Immunity test
IEC60601 test
level
Compliance level Electromagnetic
environment-guidance
Electrostatic discharge
(ESD)
IEC61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30 %.
Electrical fast
transient/burst
IEC61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage Dips, short
interruptions and
voltage
variations on power
supply input lines
IEC61000-4-11
< 5 % UT (> 95 %
dip in UT) for
0.5 cycle.
40 % UT (60 % dip
in UT) for 5 cycles.
70 % UT (30 % dip
in UT) for
25 cycles.
< 5 % UT (> 95 %
dip in UT) for
5 seconds.
< 5 % UT (> 95 %
dip in UT) for
0.5 cycle.
40 % UT (60 % dip
in UT) for 5 cycles.
70 % UT (30 % dip
in UT) for
25 cycles.
< 5 % UT (> 95 %
dip in UT) for
5 seconds.
Mains power quality should
be that of a typical
commercial or hospital
environment. If user of the
model CF-VCBC11U requires
continued operation during
power mains interruptions, it
is recommended that the
model CF-VCBC11U be
powered from an
uninterruptible power supply
or a battery.
Power frequency
(50/60 Hz) magnetic
field
IEC61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE
UT is the A.C. mains voltage prior to application of the test level
Read Me First