![Panasonic Aicure ANUJ5014 Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/panasonic/aicure-anuj5014/aicure-anuj5014_user-manual_3654209015.webp)
Name and Function of Each Part
7
Auto Batch Status Start Mode
On the Menu screen, ensure that ALL/STAT are inverted. Touch Auto to operate from the Auto Mode
screen.
•
When the main unit is in READY status, changing the external START 1 signal from “off” to “on”
(during auto irradiation, it is always on) initiates batch irradiation of the LEDs selected for this
model.
•
When the main unit is in READY status, pressing the EMISSION button on the main unit initiates
batch irradiation of the LEDs selected for this model.
•
If any of the LEDs to be used for irradiation are currently irradiating (NO LED status), irradiation
will not begin.
•
Emergency stop during irradiation can be achieved via setting the emergency stop signal on the
main unit to “open” (normally “closed”), when the external STOP 1 signal is turned on, or when
the EMISSION button on the main unit is on.
Auto Individual Status Start Mode
On the Menu screen, select MULTI/STAT, and touch Auto to operate from the Auto Mode screen.
•
When the main unit is in READY status, changing the external START signals from “off” to “on”
(during auto irradiation, it is always on) initiates individual irradiation of the LEDs selected for this
model.
•
When the main unit is in READY status, pressing the EMISSION button on the main unit initiates
batch irradiation of the LEDs selected for this model.
•
If an LED to be used for irradiation is currently irradiating, irradiation will not begin.
•
Emergency stop during irradiation can be achieved via setting the emergency stop signal on the
main unit to “open” (normally “closed”), when the external STOP signal number corresponding to
the irradiating LED is turned on, or when the EMISSION button on the main unit is on.
Warning
Warning
Содержание Aicure ANUJ5014
Страница 8: ......
Страница 16: ...Name and Function of Each Part 8 Switching Menu Screens The menu screens are switched as follows ...
Страница 35: ...Operation via External Signals 27 ...
Страница 37: ...Operation via External Signals 29 Batch Pulse Status Time Chart Batch Status Status Time Chart ...
Страница 39: ...Operation via External Signals 31 Individual Pulse Status Time Chart Individual Status Time Chart ...
Страница 48: ...40 15 Dimension Diagrams ...