Chapter 5. Advanced Programming
60
TMO2D User’s Manual
The next setting, Tau (down), is the tau value for calculations while the measured gas concentration is decreasing.
Increasing the Tau (down) reading increases the transmitter’s response speed.
5.8 LabCal Outputs?
The
LabCal Outputs? option enables you to calibrate (or trim) the analog outputs. After you select the output (A or B)
to calibrate, the output is driven to 20 mA. You can input the actual output value (as shown by a multimeter or
connected control device) and adjust the output. The output is then driven to 4 mA, and you can make a similar
adjustment.
FACTORY SETUP MENU
Tau (down) [XX.XX]:
Use the numeric keys to enter the desired value. Press [YES] to enter the number,
and [YES] again to confirm the entry.
FACTORY SETUP MENU
Thresh (%) [XX.XX]:
Use the numeric keys to enter the desired value. Press [YES] to enter the number,
and [YES] again to confirm the entry.
FACTORY SETUP MENU
Smooth (%) [XX.XX]:
Use the numeric keys to enter the desired value. Press [YES] to enter the number,
and [YES] again to confirm the entry.
FACTORY SETUP MENU
Window (%) [XX.XX]:
Use the numeric keys to enter the desired value. Press [YES] to enter the number,
and [YES] again to confirm the entry.
FACTORY SETUP MENU
Response Parameters?
Press [NO] to proceed to the LabCal Outputs? option.
FACTORY SETUP MENU
LabCal Outputs?
Press [YES] to calibrate the analog outputs.
Output to LabCal:
[A] B done
Use the [NO] or arrow keys to scroll to the desired output. Press [YES] to confirm the
entry.
Output A 20 mA Output
Enter mA [20.00]:
Use the numeric keys to enter actual output value (as read by a multimeter or
control device). Press [YES] to enter the number, and [YES] again to confirm the
entry.
Output A 4 mA Output
Enter mA [4.00]:
Use the numeric keys to enter actual output value (as read by a multimeter or
control device). Press [YES] to enter the number, and [YES] again to confirm the
entry.
Output to LabCal:
A [B] done
Use the [NO] or arrow keys to scroll to another output to repeat the calibration
procedure. Press [YES] to confirm the entry. If you have completed calibration,
scroll to “Done” and press [YES].
FACTORY SETUP MENU
LabCal Outputs?
Press [NO] to proceed to the Manual Offset? option.
Содержание TMO2D
Страница 1: ...panametrics com Process Analyzers 910 084 D August 2021 TMO2D Digital Display User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Chapter 1 Features and Capabilities 2 TMO2D User s Manual no content intended for this page...
Страница 24: ...Chapter 2 Installation 16 TMO2D User s Manual no content intended for this page...
Страница 26: ...Chapter 2 Installation TMO2D User s Manual 18...
Страница 64: ...Chapter 4 Basic Programming 56 TMO2D User s Manual no content intended for this page...
Страница 74: ...Chapter 6 Specifications 66 TMO2D User s Manual no content intended for this page...
Страница 76: ...Appendix A Outline and Installation Drawings 68 TMO2D User s Manual no content intended for this page...
Страница 87: ...Appendix A Outline and Installation Drawings TMO2D User s Manual 79...
Страница 89: ...Appendix B Menu Maps 81 TMO2D User s Manual no content intended for this page...
Страница 90: ...Appendix B Menu Maps TMO2D Digital Display User s Manual 82 Figure 26 Main Menu Setup and Outputs...
Страница 91: ...Appendix B Menu Maps TMO2D Digital Display User s Manual 83 Figure 27 Main Menu Relays and Tests...
Страница 92: ...Appendix B Menu Maps TMO2D Digital Display User s Manual 84 Figure 28 Main Menu Calibration Menu...
Страница 94: ...Appendix B Menu Maps TMO2D Digital Display User s Manual 86 Figure 30 Factory Setup Menu...
Страница 95: ...Appendix B Menu Maps TMO2D Digital Display User s Manual 87...
Страница 99: ...76 TMO2D User s Manual no content intended for this page...
Страница 100: ......