910-198E
8/19/03
Page 2-6
Model 8500
Note:
To terminate the calibration process at any stage, press the
[2nd] [MEAS] keys.
6.
Place the target ball into a Blue Labeled Reference Thickness
Fixture, and position the fixture onto the probe. The Primary
(Top) Display on the screen will show the thickness reading
according to the gage’s default table.
(If no value appears on the thickness display, then the disk
thickness of the fixture is beyond the measurement range for the
target ball. Repeat this step using a thinner disk.)
7.
Press [ENTER] or [CAL] and the default thickness value will
appear on the Secondary (Bottom) Display.
8.
Use the [ ], [ ], [
], and [
] keys to adjust the thickness
value in the Secondary Display to match the thickness value
printed on the blue labelled fixture.
9.
Press [ENTER] or [CAL]. The gage display will change to
“Point #2” prompting the user to enter another reference point.
10. Repeat steps 6-9 to continue entering reference thickness points.
Up to eight (8) reference points may be entered.
11. Press [MEAS] when done. The gage will display the message
“Please Wait, Creating Table...”.
12. To protect the customized table from being overwritten acciden-
tally, disable the Multi-Point Cal option as follows. Press
[SETUP]. Return to the Calibration Menu and press [ENTER].
Use the [ ] and [ ] keys to scroll to the Multi-Point Cal feature,
and set Multi-Point Cal to OFF by using the [
] and [
] keys.
Press [MEAS] to return to the Measurement screen.