Chapter 4. Specifications
Chapter 4. Specifications
4.1 General
Dew Point/Frost Point Calibration Range
•
–80°C to + 10°C (–112°F to 50°F) standard with data from –110°C to + 20°C (–166°F to 68°F)
Operating/Storage Temperature
•
–20°C to 60°C (–4°F to 140°F) operating temperature
•
–20°C to 60°C (–4°F to 140°F) process temperature
•
–40°C to 70°C (–40°F to 158°F) storage temperature
Warm-Up Time
•
Meets specified accuracy within three minutes after initial power up
Calibrated Accuracy (Dew/Frost Point)
•
±3.6°F (±2°C) above -148°F (-100°C)
•
±5.4°F (±3°C) below -148°F (-100°C)
Repeatability (Dew/Frost Point)
•
±0.4°F (±0.2°C) above -148°F (-100°C)
•
±0.9°F (±0.5°C) below -148°F (-100°C)
Response Time
•
Wet up time: -80°C DP (0.5 PPMv) to -15°C DP (1500 PPMv) in under 4 minutes
•
Dry down time: Dry down response is a function of how wet the probe has become and for how long. Regardless
of its starting and final Dew Point, 63% of a dry down step change will happen in under 10 seconds.
4.2 Electrical
Power
•
Input: 12 to 28 VDC (loop-powered, customer supplied)
•
Outputs: 4 to 20 mA analog, HART and RS-485 digital
•
Output Resolution: 0.01 mA/12 bits
•
Maximum Load Resistance: RL = (PSV x 33.33) – 300, where PSV = Power Supply Voltage
Example: Given a 24 VDC Power Supply, Max. Load Resistance = (24 X 33.33) – 300 = 500
Ω
•
Cable: Shielded 4-conductor 6 ft. (2 m) and 30 ft. (10 m), standard (consult Panametrics for custom lengths)
Input Parameters for Loop-Powered Intrinsic Safety
:
Loop Power Supply
U
i
= 28 V
P
i
= 0.7 W
L
i
= 0 µH
I
i
= 100 mA
C
i
= 0 µF
RS-485 Digital Output
U
i
= 4.0 V
P
i
= 250 mW
L
i
= 0 µH
I
i
= 250 mA
C
i
= 0 µF
Содержание HydroProII
Страница 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Preface x HygroProII User s Manual no content intended for this page...
Страница 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Страница 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Страница 100: ...Appendix B HygroProII Field Service Update User Manual 88 HygroProII User s Manual no content intended for this page...
Страница 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Страница 103: ......