background image

HygroPro

II

 User's Manual

3

Chapter 1. Installation

1.3

Mounting the Transmitter

Refer to Figure 3 below and complete the steps on the next page to install the HygroPro

II

 transmitter.

Figure 3: Key HygroPro

II

 parts relevant to installation

1.

Make sure the stainless-steel shield is in place over the sensor. This shield protects the aluminum oxide sensor 

from damage during operation.

2.

Using the integral 3/4-16 straight male thread, screw the probe end of the transmitter into the process or sample 

system fitting. Make sure not to damage the threads.

Note:

A 3/4-16 to G 1/2 thread adapter is available from Panametrics.

3.

Using a 1-1/8” wrench on the probe hex nut, tighten the probe securely into the process or sample system fitting.

CAUTION!

The O-ring on the Hygro

II

RTE mount at the base of the ¾-16 straight thread should be 

checked prior to mounting the HygroPro

II

 and replaced if necessary. Failure to replace 

the O-ring when the probe is dismounted and replaced can result in hazardous gas 

leaks.

CAUTION!

If the HygroPro

II

 will be installed directly into the process line, consult Panametrics for 

proper installation instructions and precautions before proceeding.

CAUTION!

Do not apply torque to the transmitter display module to tighten the unit into its sample 

system fitting.

Содержание HydroProII

Страница 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...

Страница 2: ......

Страница 3: ...ontains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries All third party product and company names are trademarks of their respective holders Hyg...

Страница 4: ...ii no content intended for this page...

Страница 5: ...Programming 17 2 1 1 Keypad 19 2 2 Basic Setup 19 2 2 1 Calibration Reminder settings on initial power up 19 2 2 2 Selecting Measurement Parameters 20 2 2 3 Setting up the Display 22 2 2 4 Setting up...

Страница 6: ...nterface 42 A 4 1 1 Sensor Input Channels 42 A 4 2 Host interface 42 A 4 2 1 Analog Output 42 A 4 2 2 Digital Output 42 A 4 3 Local Interfaces Jumpers and Switches 42 A 4 3 1 Local Controls and Displa...

Страница 7: ...le Point 61 A 9 22 Command 161 0xA1 Read Saturation Table Point 62 A 9 23 Command 162 0xA2 Set Temperature Coefficients 62 A 9 24 Command 163 0xA3 Read Temperature Coefficients 63 A 9 25 Command 166 0...

Страница 8: ...2 To enter the password 80 B 2 2 3 Incorrect Passwords 81 B 2 2 4 Exiting without inputting a passcode 81 B 3 Using the Bootloader 81 B 3 1 The Main Menu 81 B 3 1 1 Accessing the sub menus 81 B 3 1 2...

Страница 9: ...e information that provides a deeper understanding of the situation but is not essential to the proper completion of the instructions IMPORTANT These paragraphs provide information that emphasizes ins...

Страница 10: ...cable to the auxiliary equipment Examples include safety glasses protective headgear safety shoes etc Unauthorized Operation Make sure that unauthorized personnel cannot gain access to the operation o...

Страница 11: ...nces that could impact health and the environment In order to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural resources we encourage you to u...

Страница 12: ...Preface x HygroProII User s Manual no content intended for this page...

Страница 13: ...sensors and read out electronics The HygroProII transmitter uses a state of the art Aluminum Oxide trace moisture sensor that is temperature compensated for optimal accuracy It includes a sample temp...

Страница 14: ...application engineer and observe the following guidelines The sample system should be very simple and should contain as few components as possible All or most of those components should be located do...

Страница 15: ...he threads Note A 3 4 16 to G 1 2 thread adapter is available from Panametrics 3 Using a 1 1 8 wrench on the probe hex nut tighten the probe securely into the process or sample system fitting CAUTION...

Страница 16: ...s can be done in three easy steps as described below Step 1 Connect the shielded 4 wire cable supplied with the HygroProII to the keyed bulkhead connector on the back of the transmitter using the thre...

Страница 17: ...surement signal lines and its power supply lines Follow the installation instructions in this section only for units not requiring hazardous classified area protection IMPORTANT To install the HygroPr...

Страница 18: ...running TeraTerm or other terminal emulator that can used for field upgrade of the HygroPro firmware proceed to page 9 IMPORTANT To install the HygroProII in a hazardous classified area see Hazardous...

Страница 19: ...4 1 2 Standard connection with a HART master handheld device or control system 1 Push the keyed female M8 connector end of the factory supplied cable into the mating male connector on the back of the...

Страница 20: ...n the cable over the connectors and turning it clockwise until it is tight 3 Using the flying leads at the other end of the factory supplied cable connect the transmitter to the RS 485 to USB adapter...

Страница 21: ...er s manual Note It is the installer s responsibility to follow all applicable technical regulations standards and procedures Figure 10 HygroProII Installation Control Drawing for Division 1 and Divis...

Страница 22: ...drawing required Non incendive Field Wiring and Zener Barriers For use in Class I II III Division 2 hazardous areas only Figure 11 Figure 11 HygroProII Installation Control Drawing for Non incendive...

Страница 23: ...diagram This installation uses an MTL706 Zener barrier in accordance with the second option listed above Table 3 European Union EU Standards Title Number Date Electrical apparatus for explosive gas a...

Страница 24: ...communications with a PC an isolated RS232 RS485 converter mounted in the safe area between the computer and HygroProII must be used The converter is typically powered by its own standard 24 VDC powe...

Страница 25: ...eld communication to take place between one or more HART masters and intelligent field transmitters such as the HygroProII Superimposing digital signals on the analog output makes it possible for addi...

Страница 26: ...me pair of wires and communicate with the HART master The HART devices are typically set to addresses other than 0 and the loop current mode is set to disabled In multi drop mode the devices will comm...

Страница 27: ...o that particular device by HART communication It is written as plain text but is converted into a coded tokenised form for use for greater efficiency The DD file contains the standard table view and...

Страница 28: ...t can use for an operator to find each variable or method The standard DD has three top level menus Process Variables Diagram Service and Detailed Setup Each of these menus have several sub menus prov...

Страница 29: ...unit will meet its specified Dew Point accuracy within 3 minutes As noted in the previous Chapter dry down response of the HygroRTE probe depends on the initial moisture level of the AlOx sensor If t...

Страница 30: ...Loop Signaling Enable Live mA Disable Fixed mA User Passcode Cal Reminder Set Reminder Set Date CS Table Hygro Curve Pressure Curve Point 6 Zero mV Zero Psig Point 16 Point 2 Point 1 Point 15 Constant...

Страница 31: ...ng six top level parameters can be accessed and set on the HMI without any passcode a No of Views b Display c Output d HART e Constants f About The Settings menu requires a 6 digit User passcode defau...

Страница 32: ...nce the HygroProII does not use a real time clock battery the clock needs to be set upon power up If the user chooses either 6 months or 1 year they are then required to entire the current time and da...

Страница 33: ...o choose Press Enter Again any change in this selection is confirmed by a Save Successful prompt flashing on the screen after pressing Enter Use the arrow keys to highlight Temp and press Enter to set...

Страница 34: ...ase note each measurement in HygroProII has a pre determined range and resolution and the number of digits after the decimal point is fixed 2 2 3 Setting up the Display To set up the display complete...

Страница 35: ...n will disappear from the display After pressing Escape to access the Main Menu use the right arrow key to scroll to Output and press Enter Select the Measurement option and press Enter Use the arrow...

Страница 36: ...and press Enter The screen shown on the left appears Enter the Zero and Span values for the range using the same procedure as above In the Output menu select Trim and press Enter The following screen...

Страница 37: ...bar could be selected if the desired measurement output is the Equivalent DP 70 bar CAUTION Please be careful to select sample pressure and temperature settings to the best option for the chosen hygro...

Страница 38: ...nd press Enter It is advisable to start at the top and check every single calibration point in order to ensure the probe calibration curve has been entered properly The next screen shows the FH indepe...

Страница 39: ...under the Service menu see Chapter 3 The end user can adjust all other device parameters shown in the Menu Map in Figure 16 Since the probe calibration data are protected by the User passcode we stron...

Страница 40: ...w keys to select Enable Disable HART and press Enter The next screen shown on the left enables you to either disable default is enabled or enable HART on the 4 20 mA output Press Enter In a multi drop...

Страница 41: ...tions choose the About option on the Main Menu and press Enter to access the About Menu screen Use the Up and Down arrow keys to select About Display and press Enter The next screen shown on the left...

Страница 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...

Страница 43: ...ntermediate screen Select the desired Service menu option and press Enter Then follow the on screen instructions After unlocking the display if required and pressing Escape use the arrow keys to scrol...

Страница 44: ...AlOx sensor on the probe After completing the above step respond to the Are you sure question Be sure to respond Yes if you wish to save the new information Please note your input has not been accepte...

Страница 45: ...accomplish this either return the RTE to Panametrics for recalibration or install a new RTE The HygroProII electronics will automatically read and store the calibration data whenever a new or recalib...

Страница 46: ...g the Probe from the Transmitter The HygroII RTE probe is easy to remove from the dismounted transmitter by loosening the two shoulder screws and disconnecting the internal M8 connector cable as shown...

Страница 47: ...or cleaned probe back in the transmitter see Figure 19 below This procedure requires completing the following steps 1 Ensure the two shoulder screws at the bottom of the display head are fully loosene...

Страница 48: ...mitter 2 Compare the two ambient air readings If the new ambient air reading is within 2 C 3 6 F of the first reading the cleaned probe is properly calibrated and normal operation may be resumed 3 If...

Страница 49: ...ime Wet up time 80 C DP 0 5 PPMv to 15 C DP 1500 PPMv in under 4 minutes Dry down time Dry down response is a function of how wet the probe has become and for how long Regardless of its starting and f...

Страница 50: ...librated against known moisture concentrations traceable to NIST Calibration Interval Sensor recalibration at Panametrics is recommended every six to twelve months depending on the application Flow Ra...

Страница 51: ...000 psig 35 to 345 bar Note PSIG ranges are based on a constant pressure with the value provided at the time of order placement Accuracy 1 of full scale FS Warm up Time Meets specified accuracy within...

Страница 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...

Страница 53: ...ication is designed to be a technical reference for HART capable host application developers system integrators and knowledgeable end users It also provides functional specifications e g commands enum...

Страница 54: ...A 4 1 Process Interface A 4 1 1 Sensor Input Channels A single removable transducer element moisture probe is connected to the device Refer to the manuals for connection instructions A 4 2 Host inter...

Страница 55: ...ees C 0 0 Moisture Dew Point Degrees F 1 0 Moisture Parts Per Million Volume PPMv 2 0 Moisture Parts Per Million Volume PPBv 3 0 Moisture Parts Per Million Volume Natural Gas PPMvNG 4 0 Moisture mg m3...

Страница 56: ...mmand 48 Command 48 returns 25 bytes of data with the following status information HART Additional Device Status Byte Bit Error Description Class Device Status Bits Set 0 0 Reserved Warning 4 7 1 No P...

Страница 57: ...0 Byte 10 is the Analog Channel Saturated Byte if the Loop Current is saturated the lowest bit will read 1 otherwise all bits are 0 Byte 13 is the Analog Channel Fixed Byte if the loop current is fixe...

Страница 58: ...he Command 0 response 12 Read Message Reads the Message contained within the meter 13 Read Tag Descriptor Date Reads the Tag Descriptor and Date contained within the meter 14 Read Primary Variable Tra...

Страница 59: ...mmand 148 0x94 Set Temperature Constant Command 149 0x95 Read Temperature Constant Command 150 0x96 Set Dew Point Offset Command 151 0x97 Read Dew Point Offset Command 152 0x98 Set Hygro Curve Number...

Страница 60: ...ro value is 4 1 mA then the correction entered should be 0 1 mA to achieve a result of 4 mA Request Data Bytes Response Data Bytes Command Specific Response Codes Byte Format Description 0 Unsigned 8...

Страница 61: ...nalog output or fixed 20 mA for trim purposes 1 4 Float mA correction value Byte Format Description 0 Unsigned 8 Enumeration 0 Fixed 20 mA trim 1 Live mA Set for live analog output or fixed 20 mA for...

Страница 62: ...utput or fixed mA value 1 4 Float Loop Current Percentage units of percent Byte Format Description 0 Unsigned 8 Enumeration 0 Fixed percent test 1 Live mA Sets live analog output or fixed mA value 1 4...

Страница 63: ...and Lower Range Values Unit Code 2 5 Float Upper Range Value 6 9 Float Lower Range Value Code Class Description 0 Success No Command Specific Errors 1 Undefined 2 Error Invalid Selection 3 4 Undefined...

Страница 64: ...n the Special span value and special zero value will determine the endpoint mA values instead of 4 mA and 20 mA Request Data Bytes Response Data Bytes 32 Error Busy 33 127 Undefined Byte Format Descri...

Страница 65: ...mand Specific Errors 1 Undefined 2 Error Invalid Selection 3 4 Undefined 5 Error Too Few Data Bytes Received 6 Error Device Specific Command Error 7 Error In Write Protect Mode 8 15 Undefined 16 Error...

Страница 66: ...to be used in place of a live measurement First byte is an enumeration 0 temperature constant active 1 live measurement instead of constant Request Data Bytes Response Data Bytes 16 Error Access Rest...

Страница 67: ...x96 Set Dew Point Offset This command will set the Dew Point Offset adjustment Request Data Bytes Code Class Description 0 Success No Command Specific Errors 1 4 Undefined 5 Error Too Few Data Bytes R...

Страница 68: ...of points for the Hygro Curve Request Data Bytes Byte Format Description 0 3 Float Dew Point offset adjustment Code Class Description 0 Success No Command Specific Errors 1 4 Undefined 5 Error Too Few...

Страница 69: ...Point Request Data Bytes Byte Format Description 0 Unsigned 8 Number of Hygro Calibration Points Code Class Description 0 Success No Command Specific Errors 1 Error Number of points out of range 2 4...

Страница 70: ...of point Code Class Description 0 Success No Command Specific Errors 1 Error Index Out of Bounds 2 127 Undefined Byte Format Description 0 Unsigned 8 Index of Hygro curve point 1 2 Signed 16 Dew Poin...

Страница 71: ...tes Byte Format Description 0 3 Float Low End mV 4 7 Float Low End PSIg 8 11 Float High End mV 12 15 Float High End PSIg Byte Format Description 0 3 Float Low End mV 4 7 Float Low End PSIg 8 11 Float...

Страница 72: ...uration table points Request Data Bytes 8 11 Float High End mV 12 15 Float High End PSIg Code Class Description 0 Success No Command Specific Errors 1 127 Undefined Byte Format Description 0 Unsigned...

Страница 73: ...iption 0 Success No Command Specific Errors 1 127 Undefined Byte Format Description 0 Unsigned 8 Index of saturation table point 1 4 Float Temperature degC 5 8 Float Saturation Constant Byte Format De...

Страница 74: ...ecific Response Codes Byte Format Description 0 Unsigned 8 Index of Saturation point Byte Format Description 0 Unsigned 8 Index of Saturation Table point 1 4 Float Tempearture deg C 5 8 Float Saturati...

Страница 75: ...lement Request Data Bytes Response Data Bytes 5 Error Too Few Data Bytes Received 6 Error Device Specific Command Error 7 Error In Write Protect Mode 8 15 Undefined 16 Error Access Restricted 8 127 Un...

Страница 76: ...ent Set the Hygro Calibration Curve index for the reading of an array element Request Data Bytes Response Data Bytes Command Specific Response Codes Code Class Description 0 Success No Command Specifi...

Страница 77: ...d Span Values When the loop measurement type is set to Special then the Special span value and special zero value will determine the endpoint mA values instead of 4 mA and 20 mA Request Data Bytes Res...

Страница 78: ...the zero value is 20 1 mA then the correction entered should be 0 1 mA to achieve a result of 20 mA Request Data Bytes 6 Error Device Specific Command Error 7 Error In Write Protect Mode 8 15 Undefine...

Страница 79: ...Live mA Read for live analog output or fixed 4 mA for trim purposes 1 4 Float mA correction Value Code Class Description 0 Success No Command Specific Errors 1 4 Undefined 5 Error Too Few Data Bytes...

Страница 80: ...9 Error Incorrect Loop Current Mode or Value 10 Undefined 11 Error Loop Current Not Active 12 15 Undefined 16 Error Access Restricted 17 31 Undefined 32 Error Busy 33 127 Undefined Command Function P...

Страница 81: ...perature Constant 150 Set Dew Point Offset 151 Read Dew Point Offset 152 Set Hygro Curve Number of Points 153 Read Hygro Curve Number of Points 154 Set Hygro Curve Point 155 Read Hygro Curve Point 156...

Страница 82: ...anced user can change the General user password and Advanced user password Request Data Bytes Response Data Bytes Byte Format Description 0 Unsigned 8 Enum User level 0 None 2 Operator User 3 Admin Us...

Страница 83: ...meters will not be available after the HygroProII reboots automatically until the password is right and this command is sent to the HygroProII in 10 minutes After the HygroProII reboots automatically...

Страница 84: ...nt 148 Set Temperature Constant 149 Read Temperature Constant 150 Set Dew Point Offset 151 Read Dew Point Offset 152 Set Hygro Curve Number of Points 153 Read Hygro Curve Number of Points 154 Set Hygr...

Страница 85: ...es Command Specific Response Codes Byte Format Description None Byte Format Description None Code Class Description 0 Success No Command Specific Errors 1 5 Undefined 6 Error Device Specific Command E...

Страница 86: ...Types 240 Moisture 240 Dew Point Degrees C 241 Dew Point Degrees F 243 Parts Per Million Volume PPMv 244 Parts Per Million Volume PPBv 245 Parts Per Million Volume Natural Gas PPMvNG 248 mg m3 249 g m...

Страница 87: ...7 Long Messages The largest data field used is in the response to Command 183 21 bytes including the two status bytes A 11 8 Non Volatile Memory EEPROM is used to hold the device s configuration para...

Страница 88: ...mon practice commands 6 Number of device specific commands 58 Bits of additional device status 11 Alternative operating modes No Burst mode No Write protection Yes Parameter Default value Lower Range...

Страница 89: ...M8 connector usually White b Connect the RS485 Data B to the wire corresponding to Pin 4 of the M8 connector Usually Black Figure 21 Pin Out for 6 Pin Female M8 Connector 2 Connect the 5 Wire cable t...

Страница 90: ...lect Serial b Select the USB to RS485 adapter from the drop down menu It will be labeled USB Serial Port Figure 22 Select Serial Connection 4 Set the correct settings for the serial terminal a From th...

Страница 91: ...correct font and text size Optional for readability a From the menu at the top of the window select Setup Font b Set the font to Arial c Set the font style to Regular d Set the size to 12 e NOTE These...

Страница 92: ...or the HygroProII Bootloader is generated using the unique Device ID Number The Bootloader password is separate from the Service passcode This Device ID number is independent from the device Serial Nu...

Страница 93: ...un B 3 Using the Bootloader B 3 1 The Main Menu The Main Menu is where you can access all the features of the bootloader Figure 26 Main Menu B 3 1 1 Accessing the sub menus 1 To access the sub menu ty...

Страница 94: ...Details Menu will show the version numbers for both the installed bootloader and the installed instrument program If the bootloader cannot verify that the firmware is genuine or if the firmware is cor...

Страница 95: ...ils Menu and return to the Main Menu B 3 3 Download and Install New Firmware Menu This is where you can download and install both new bootloader and instrument program firmware The device will require...

Страница 96: ...stall new bootloader firmware b Press 2 to install new Instrument program firmware 4 The menu below will be shown a NOTE IMPORTANT It is important that the device does not lose power from this point u...

Страница 97: ...YMODEM Send a NOTE If you wish to exit this menu you must press a twice in a row This is a special command required to interrupt the file transfer Figure 31 Firmware Download Menu 6 A file browser win...

Страница 98: ...will be shown the file details Figure 33 Downloaded Firmware Details 12 The bootloader will attempt to verify the downloaded firmware binary a If the instrument program firmware is valid it will be i...

Страница 99: ...ation data stored in the flash memory of the device For the HygroProII Probe this includes all calibrations calibration date and serial numbers For the HygroProII Display this includes all 4 20mA outp...

Страница 100: ...Appendix B HygroProII Field Service Update User Manual 88 HygroProII User s Manual no content intended for this page...

Страница 101: ...particular purpose and warranties arising from course of dealing or usage or trade Return Policy If a Panametrics instrument malfunctions within the warranty period the following procedure must be com...

Страница 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...

Страница 103: ......

Страница 104: ...duct and company names are trademarks of their respective holders 910 331 A Customer Support Centers U S A The Boston Center 1100 Technology Park Drive Billerica MA 01821 U S A Tel 800 833 9438 toll f...

Отзывы: