Chapter 2. Error Codes and Screen Messages
18
DigitalFlow™ GF868 Service Manual (1 and 2-Channel)
2.1.1
E0: No Error
Problem:
No error condition currently exists.
Cause:
This message appears briefly to confirm that the response to another error message has corrected the
problem.
Action:
No action is required.
2.1.2
E1: Low Signal
Problem:
Poor ultrasonic signal strength or the signal exceeds the limits entered via the
User Program.
Cause:
Poor signal strength may be caused by a defective cable, a flowcell problem, a defective transducer or a
problem in the electronics console. A signal that exceeds the programmed limits is probably caused by
the entry of an improper value in the SETUP SIGNL submenu of the User Program.
Action:
Using the procedures in Chapter 3,
Diagnostics, check the components listed above. Also, check the
value programmed into the SETUP SIGNL submenu, as described in the
Operator’s Manual.
2.1.3
E2: Sound Speed Error
Problem:
The receive signal is detected too close to the open or close of the receive window.
Cause:
Incorrect programming of the path length. Loss of signal, so that noise at the start of the window is
triggering the measurement.
Action:
Verify programming of the path length. Check for loss of signal.
2.1.4
E3: Velocity Range
Problem:
The velocity exceeds the limits programmed in the SETUP submenu of the
User Program.
Cause:
This error may be caused by the entry of improper programming data or by poor flow conditions and/or
excessive turbulence.
Action:
Make sure the actual flow rate is within the programmed limits. See the
Programming Manual for details.
Refer to Chapter 3,
Diagnostics, to correct any flowcell and/or transducer problems.
2.1.5
E4: Signal Quality
Problem:
The signal quality is outside the limits programmed in the SETUP submenu of the
User Program.
Cause:
The peak of the upstream or downstream correlation signals has fallen below the correlation peak limit,
as set in the
SIGNL
submenu. This may be caused by a flowcell or electrical problem.
Action:
Check for sources of electrical interference and verify the integrity of the electronics console by
temporarily substituting a test flowcell that is known to be good. Check the transducers and relocate
them, if necessary. See Chapter 3,
Diagnostics, for instructions.
Содержание GF868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 24: ...Chapter 1 Calibration DigitalFlow GF868 Service Manual 1 and 2 Channel 16...
Страница 36: ...Chapter 3 Diagnostics 28 DigitalFlow GF868 Service Manual 1 and 2 Channel no content intended for this page...
Страница 62: ...Warranty 60 DigitalFlow GF868 Service Manual 1 and 2 Channel...
Страница 66: ...64 DigitalFlow GF868 Service Manual 1 and 2 Channel no content intended for this page...
Страница 67: ......