Chapter 1. Calibration
12
DigitalFlow™ GF868 Service Manual (1 and 2-Channel)
1.6 Testing the Alarm Relays (cont.)
1.6.1
Accessing the Calibration Menu
1.
Press the
[CAL]
key to enter the
Calibration Program.
2.
Press
[Fx]
to calibrate
the desired slot
. (The option bar will include a slot listing for each installed option card.)
3.
Press
[F1]-[F3]
to select alarm relay A, B or C, respectively.
Note:
The procedure for testing
Alarm Relay A
is identical to that for testing
Alarm Relays B and C
. However, make
sure that the ohmmeter is connected to the desired normally-open or normally-closed contact of the
currently selected relay. See Figure 4 on page 11 for the correct pin numbers on the option card terminal
block.
4.
Pressing
[F1]
(CLOSE)
should yield an ohmmeter reading of about zero. Pressing
[F2]
(OPEN)
should yield an
infinite ohmmeter reading.
5.
Press
[F3]
to
EXIT
.
6.
Repeat the above procedure until both the normally-open and normally-closed contacts for all three alarm
relays have been tested. Then proceed to
Procedure Options
below.
1.6.1.1
Procedure Options
You have completed testing the alarm relays. Do one of the following:
• To calibrate/test additional slots, press
[F4] (EXIT)
and proceed to the appropriate section.
• To leave the Calibration Menu, press
[F4] (EXIT)
twice.
Содержание GF868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 24: ...Chapter 1 Calibration DigitalFlow GF868 Service Manual 1 and 2 Channel 16...
Страница 36: ...Chapter 3 Diagnostics 28 DigitalFlow GF868 Service Manual 1 and 2 Channel no content intended for this page...
Страница 62: ...Warranty 60 DigitalFlow GF868 Service Manual 1 and 2 Channel...
Страница 66: ...64 DigitalFlow GF868 Service Manual 1 and 2 Channel no content intended for this page...
Страница 67: ......