52
ID08435 - Rev. 1.2 - 19/04/2023
8. SANITATION
8.1 Sanitizing products
ATTENTION! Sanitizing agents are corrosive.
The best sanitizing agents are the most commonly used in industry. Follow the instructions
provided by the manufacturer for the specific application during use. If possible, ask the
manufacturer for guarantees on the degree of corrosivity of the solutions used.
Any change to these features may damage the item.
It’s very important to follow the specifics of concentration, temperature and reaction
time. Any change to these features may damage the device. During sanitization steps
only use:
▪
cold mineral water;
▪
hot water max. 95 C;
▪
alkaline solutions max. 80 C;
▪
acid solutions;
▪
disinfectant solutions.
Do not use sulfuric acids or mineral acids such as HCl, H SO , HNO and H SO .
2
4
3
2
3
8.2 Sanificazione con prodotti contenenti alogeni
WARNING: Do not use halogen-containing products during closed circuit sterilization, the
bed may be damaged.
If used incorrectly these products can corrode steel especially if the pH is low. Perform
accurate checks before using these solutions.
If the device is to be sanitised using halogen (e.g. chlorine) sanitising products, the
following requirements shall be followed:
▪
the pH must be above 10;
▪
the temperature must not exceed 40 ºC;
▪
the solution must not stay in contact with the bed for more than 20 min.;
▪
a concentration of up to 50 ppm of active chlorine shall be used;
▪
after sanitizing perform a generous rinse with water.
8.3 Sanitizing intervals
The sanitization intervals are defined by the user, according to the needs, taking into
account the indications in this manual and those in the sanitizing products used.
8.4 Automatic sanitization
The automatic sanitization (autoclave) is defined by the customer, according to the
needs, taking into account the indications in this manual and those reported by the
sanitizing products used.
8.5 Manual sanitization
The manual sanitization will be defined by the customer, according to the needs, taking into
account the indications in this manual, and those reported by the sanitizing products used.
ATTENTION! Always check the safety data sheets of the materials used for
sanitization. In case of contact/ inhalation and/ or ingestion, follow the prescriptions
indicated in the prescribed sheets.