8
EmStat3-MUX16 for OEM
EmStat3-MUX16 for OEM
Conf. 2: Sensor array with up to 16 working electrodes all sharing one reference and one counter
electrode:
Jumper to be placed: J2
When J3 is also placed, the potential is applied to all working electrodes continuously.
Use pin 1, 5, 9, 13, 17, 21, 25 or 29 to connect RE
Use pin 3, 7,11, 15,18, 23, 27 or 31 to connect CE
Conf. 3: Sensor array with up to 16 working electrodes all sharing one combined
reference/counter
electrode:
Jumpers placed are: J1 and J2
When J3 is also placed, the potential is applied to all working electrodes continuously.
Use pin 1, 3, 5….. to 31 to connect RE/CE
*
IMPORTANT REMARKS
It is not possible to apply a potential simultaneously to more than one sensor or cell each with three
electrodes. This requires a multipotentiostat, one potentiostat for each channel. This is however possible
with two electrode sensors or cells, so when combined counter and reference electrodes are applied.
The sensor configuration with 16 WE’s all having their own reference and separate counter electrode is
NOT applicable with the MUX16 multiplexer.