2.3 

Safety Messages

  

 

DANGER! 

Risk of injury to personnel and/or damage to equipment!  

Before using the iCELLis 500+ control system, ensure you read and understand the safety instructions 
to prevent any damage to the equipment or any physical injury. Pall shall not be held responsible for 
any damage that may result from misuse. Misuse includes any use other than that designed for the 
iCELLis 500+ bioreactor and control system. Keep this guide handy for later reference. In the event 
you have additional questions, please contact Pall. 

 

DANGER! 

Risk of injury to personnel and/or damage to equipment!  

Use appropriate personal protective equipment (PPE) when moving and using the iCELLis 500+ 
control system. 

 

DANGER! 

Biological hazard! 

The biological hazard is a function of the process performed and is under the responsibility of the user. 
Appropriate safety risk analysis, depending on the biological material used, should be performed by 
the user. Appropriate personal protective equipment must be used. 

 

WARNING! 

Risk of injury to personnel and/or damage to equipment!  

The iCELLis 500+ control system must be operated by qualified people within a safe environment only. 
This equipment is expected to be installed in a controlled environment. Do not use the equipment in 
hazardous locations, such as Atmosphere Explosive (ATEX) Environmental operating conditions: 
18 - 25 °C, 10 - 70 % relative humidity non-condensing. When not in use, the iCELLis 500+ control 
system must be stored in temperatures ranging from 4 ºC - 40 ºC, 80% maximum relative humidity 
non-condensing. Keep the iCELLis 500+ control system in a dry place. 

 

ALERT! 

Risk of damage to equipment!  

This equipment has been designed for indoor use only.  

 

DANGER! 

Risk of injury to personnel and/or damage to equipment!  

Before installing the iCELLis 500+ control system, ensure that the floor can support the weight of the 
system, the ancillary equipment and their content (such as cell culture media). 

 

ALERT! 

Risk of damage to equipment! 

Never use the equipment in a manner not specified by or clearly prohibited by the manufacturer. 
Such inappropriate use may impair the equipment’s built-in protections and would void the warranty. 

 

ALERT! 

Risk of damage to equipment! 

Do not use the iCELLis 500+ control system in an incubator. 

 

ALERT! 

Risk of damage to equipment!  

Never position the equipment so that it is difficult to access the machine main power switches, gas inlet 
ports and emergency stop buttons. A minimum clearance of 650 mm around the system (front, rear, 
left and ride sides) is needed. 

 

WARNING! 

Risk of injury to personnel and/or damage to equipment!  

Refer to the operating manual of the Temperature Control Unit (TCU) provided within the 
documentation of the iCELLis 500+ control system. 
Make sure you read and understand all instructions and safety precautions listed in the operating 
manual of the TCU before installing or operating your unit. 

 

WARNING! 

Risk of injury to personnel and/or damage to equipment!  

Place the TCU on an even surface with a base made of nonflammable material. 

 

WARNING! 

HOT/COLD surfaces! Risk of injury to personnel and/or damage to equipment! 

Always empty the bath of the TCU before moving the unit. Do not drain the bath fluid while it is hot! 
Check temperature of bath fluid prior to draining (e.g. by switching the unit on for a short moment). 

 

WARNING! 

Bended tubing! Risk of injury to personnel and/or damage to equipment!  

Avoid sharp bends in the fluid tubing of the TCU and maintain a sufficient distance from surrounding 
walls. Make sure that the tubing is securely attached. 

 

CAUTION! 

Risk of injury to personnel and/or damage to equipment! 

When the TCU, media pre-heater or the filter heater are active the surfaces and tubes can be hot. 
Therefore, exercise particular caution when handling. 

 

ALERT! 

Risk of damage to equipment!  

Do not disconnect the communication cable between the iCELLis 500+ control system and the TCU 

Содержание iCELLis 500+

Страница 1: ...lution SM 1 1 1 1 Instructions for Use Part A Hardware USD3310 iCELLis 500 Bioreactor Control System Part Number ICL500CSSSIP Instructions For Use Part A Hardware USD 3310 iCELLis 500 control system P...

Страница 2: ...___________________22 5 6 Fitting Communication and Sensor Cables to the iCELLis 500 Control System _________23 5 6 1 Connecting the Double Jacket Pressure Sensor Cable _______________________________...

Страница 3: ...osal __________________________________________________________ 45 10 Technical Assistance _______________________________________________ 46 Appendix A Abbreviations and Symbols _____________________...

Страница 4: ...used in accordance with instructions detailed in this manual Utilization must be carried out only by authorized persons trained to the iCELLis 500 control system operation Operation of the equipment m...

Страница 5: ...ol notifies a risk of electric shock Crush WARNING icon This symbol notifies a risk of a crushing event Hot Surface WARNING and CAUTION icon This symbol notifies a risk of burn Magnetic field DANGER i...

Страница 6: ...CAUTION wiring color differs from the NFPA 79 Standards The electric wire colour used in the iCELLis 500 control system is listed on the label and conformed to the IEC 60445 2 2 Hazard Level Hazard ic...

Страница 7: ...t the weight of the system the ancillary equipment and their content such as cell culture media ALERT Risk of damage to equipment Never use the equipment in a manner not specified by or clearly prohib...

Страница 8: ...or damage to equipment User is responsible for performing assessment on the risks linked to liquids used in the iCELLis 500 control system and bioreactor Appropriate personal protective equipment mus...

Страница 9: ...Lis 500 control system is powered ALERT Risk of damage to equipment Only exhaust filter heater provided by Pall should be used with the iCELLis 500 control system ALERT Risk of damage to equipment The...

Страница 10: ...not change with the fixed bed volume and offers a maximum working volume of 74 L Further media volume can be added through perfusion recirculation Hence the same docking and controlling station is us...

Страница 11: ...ps probe connections probe transmitters electrical connections emergency stop buttons and the touch screen panel An external TCU for bioreactor temperature control The bioreactor is placed on the dock...

Страница 12: ...overview of the iCELLis 500 control system components in front rear right and left views are presented in Figure 3 Figure 4 and Figure 5 Optional High Flow peristaltic pumps Touch screen panel Middle...

Страница 13: ...Page 13 of 48 Figure 3 iCELLis 500 control system front view Figure 4 iCELLis 500 control system rear view...

Страница 14: ...Page 14 of 48 Figure 5 iCELLis 500 control system right and left views...

Страница 15: ...isual alarm and tubing holders 2122 mm with visual alarm and tubing holders Width 955 mm excluding optional high flow pump 1038 mm with optional high flow pump Length 1609 mm Dry Weight without biorea...

Страница 16: ...nd weight Table 3 Weight and dimensions of iCELLis 500 control system crates iCELLis 500 Control System Accessories and Documentation Crate weight 850 kg 300 kg Crate dimensions H x W x L 2135 x 1120...

Страница 17: ...ted unit b Removing crate fixing clips c Lifting side panel d Removing side panel e Side panel removed f Unit with double plastic wrap g Unit with external wrap removed h Unit unwrapped and lifting po...

Страница 18: ...nly the bottom assembly must be used to push or pull the unit Do not push or pull from the docking station Please do not move the equipment in any other way than instructed Unlocking Load Cells ALERT...

Страница 19: ...the wheels should be turned and feet lowered to lock the casters Figure 8 When performing this step a spirit level should be used and placed on the bioreactor docking station to ensure the hardware is...

Страница 20: ...lug the system from the AC power shut down gas supply disconnect gas lines remove the bioreactor remove the manifolds and all liquids The high speed pumps are packed in the crate for accessories and d...

Страница 21: ...re 10a b Connect the base scale at the back of the iCELLis 500 control system to the scale connector Figure 10c Place the Base tray on the scale Figure 10d Level the scale by turning the blue feet of...

Страница 22: ...e 11 Item number Description Quantity 1 Seal washer M16 4 2 Left tube holding channel 2 1 3 Left tube holding channel assembly 1 4 Hex drive button head screwed in 7380 Stainless steel A2 M16 x 30 mm...

Страница 23: ...gure 12 Double jacket pressure sensor cable Connect the cable to the connector labeled DJ PRESSURE SENSOR located on the front of the iCELLis 500 control system Figure 13a Secure the connector by turn...

Страница 24: ...d BIOREACTOR PRESSURE SENSOR located on the front of the iCELLis 500 control system Figure 15a Secure the connector by turning the metallic part of the connector Figure 15b a b Figure 15 Bioreactor pr...

Страница 25: ...tion Figure 16 T1 grey cable and T2 white cable temperature probe Connect the T1 temperature probe cable to the connector labeled T1 located on the front of the iCELLis 500 control system Figure 17a S...

Страница 26: ...nd pH1 short ones and after DO2 and pH2 long ones the packed bed are packed in the crate of the accessories and documentation Figure 18 Optical fibers before short and after long fix bed Connect the o...

Страница 27: ...b c d e Figure 19 Connection of optical fibers before and after fixed bed a Optical connectors b Connection to the DO2 connector c Secured connection d Connection to the pH2 connector e All optical f...

Страница 28: ...wer connector of the filter heater to the connector labeled POWER on the right side of the iCELLis 500 control system Figure 21a and b Secure the connector by turning the metallic part of the connecto...

Страница 29: ...of 48 a c b d e Figure 21 Connectors of the filter heater a b Connection of the power connector c Secured power connection d Connection of the temperature sensor e Secured connection of the temperatur...

Страница 30: ...upper part of the bag pre heater install the metric hex screws with a metric hex key of 5 mm locking the bottom part of the bag pre heater on the iCELLis 500 control system Figure 23b c a b c Figure...

Страница 31: ...in the double jacket of the bioreactor DANGER Risk of injury to personnel and or damage to equipment The ingress protection of the TCU is IP 21 please refer TCU user manual for cleaning instructions W...

Страница 32: ...30Vac 50Hz and the other must be supplied with 120Vac 60Hz The configuration of the TCU must be specified according to the client requirement during the purchasing of the iCELLis 500 control system Th...

Страница 33: ...c d Take out the Profibus cable from crate and connect one end of the cable to right side of the TCU and other end to back of the iCELLis 500 control system tighten the connector on the TCU using a fl...

Страница 34: ...nd touch panel e Connection of Profibus connector on TCU f Other end of Profibus cable on iCELLis 500 bioreactor g TCU rear side image h Power cable i Connecting power on rear side of TCU j Barbed con...

Страница 35: ...er supply cable Only use the cable supplied with the unit Insure that the iCELLis 500 control system is connected to suitable overcurrent protection device Protective grounding for the system is provi...

Страница 36: ...expert advice on the design of any system used to supply gases to the iCELLis 500 control system RELEVANT ONLY WHEN USING OXYGEN In the unlikely event of a tubing or connection failure oxygen may buil...

Страница 37: ...n independent pressure regulator All materials in contact with the 4 gases are made of stainless steel or USP 88 Class VI material Figure 28 Gas supply inlets of iCELLis 500 control system Gas input p...

Страница 38: ...SO9001 2008 The product is made in a manufacturing facility where the Quality Management System is ISO9001 2015 certified Waste Electrical and Electronic Equipment WEEE The hardware meets the requirem...

Страница 39: ...ted impeller in the iCELLis disposable bioreactor The impeller is magnetically driven The impeller speed can be set via the iCELLis 500 software between 0 and 450 rpm Temperature control Bioreactor in...

Страница 40: ...Compatible with FDA 21 CFR Part 11 and Eudralex Annex 11 Automation design Developed and tested in accordance with GAMP5 guidelines Data export Network communication Integrated OPC server for real tim...

Страница 41: ...the fixed bed in mS cm 2 0 40 0 3 0 mS cm AT_06 Biomass of the fixed bed in pF cm 5 400 3 0 pF cm FT_01 CO2 flow transmitter 0 00 1 50 0 1 NL min FT_02 N2 flow transmitter 0 00 1 50 0 1 NL min FT_03...

Страница 42: ...450 5 0 rpm ST_07 Speed fill pump optional Watson Marlow 630BpN 0 160 2 0 rpm ST_08 Speed drain pump optional Watson Marlow 630BpN 0 160 2 0 rpm 6 3 Temperature Control Unit Specifications Descriptio...

Страница 43: ...olecular weight alcohol especially isopropyl alcohol to decontaminate the exterior of the iCELLis 500 bioreactor may in circumstances where significant stress repetitive bending and twisting is applie...

Страница 44: ...any warranties Please contact Pall See Section 10 In any correspondence with Pall the model manufacturing part and serial numbers of the hardware should be recorded for reference It is recommended to...

Страница 45: ...ical and electronic equipment WEEE According to these regulations any device supplied after August 13 2005 in the business to business sphere to which this product is assigned may no longer be dispose...

Страница 46: ...bor Park Drive Port Washington NY 11050 USA 1 800 717 7255 toll free USA 1 516 484 5400 phone biopharm pall com e mail European Headquarters Avenue de Tivoli 3 1700 Fribourg Switzerland 41 0 26 350 53...

Страница 47: ...I Human Machine Interface Hz Hertz IPA Isopropyl Alcohol m mm Meter millimeter MFC Mass Flow Controller OD Outer Diameter P ID Piping and Instrumentation Diagram PET Polyethylene terephthalate PETG Po...

Страница 48: ...e time of publication Product data may be subject to change without notice For current information consult your local Pall distributor or contact Pall directly Asia Pacific Headquarters Singapore 65 6...

Отзывы: