Message Error
To solve
Error 100
Measurement The reading is above 1050 NTU.
Error 101
Incorrect
Standard
The standard inserted is either
the incorrect value or is no
longer within an acceptable
tolerance for calibration. Use the
correct standard or replace the
calibration set.
Error 107
Light
Detection
The sample tube cap is not in place
or the sample tube is not fully
inserted. Replace the cap and read
again or fully insert the sample tube.
If the error persists, set the zero
reference level by pressing the ‘Ref
Reset’ whilst a capped tube is in
the holder
Error 110
Battery Low
Instrument has insufficient power to
perform reading. Menu functions still
available.
Error 111
Battery
Critical
Insufficient battery to power
instrument. Instrument will save
data and shut down.
The Turbimeter Plus has a two year warranty.
Calibration standards should not be used after their stated
expiry date. Replacement sets are available with part code
PTC 092.