5
Test 2
TOTAL CHLORINE
Option 2: Range 0 - 10.0 mg/l (ppm)
Colourless - Red
1 Carry out this test on the solution remaining from the Free Chlorine test.
2 If any shock treatment chemicals have been added to the pool, add one DPD
Oxystop tablet, crush and mix to dissolve. This will prevent a response caused
by the shock treatment chemicals.
3 Add one DPD-XT tablet, crush and mix to dissolve.
4 Stand for two minutes.
5 Take photometer reading. Ensure display shows the Chlorine-Total/10 test is
selected.
To obtain COMBINED CHLORINE residual subtract Free Chlorine result from Total
Chlorine result :-
ie Combined Chlorine = Total Chlorine - Free Chlorine
Note: To assess the level of shock dose chemicals in the pool, request supplementary
instructions Phot 7.2.AUTO.
Test 3
BROMINE
Range 0 - 6.00 mg/l (ppm)
Colourless - Red
1 Rinse test tube with sample leaving two to three drops in the tube.
2 Add one DPD No 1 tablet, crush tablet and then fill the test tube with sample to
the 10 ml mark. Mix to dissolve tablet and ensure any remaining particles have
settled.
3 Take photometer reading immediately, as result may drift on standing.
Note: To separate chlorine and bromine residuals, request supplementary
instructions Phot 5.1.AUTO.
Test 4
OZONE
Range 0 - 2.00 mg/l (ppm)
Colourless - Red
1 Rinse test tube with sample leaving two to three drops in the tube.
2 Add one DPD No 4 tablet, crush tablet and then fill the test tube with sample to
the 10 ml mark. Mix to dissolve tablet and ensure any remaining particles have
settled.
3 Take photometer reading immediately.
This test also responds to chlorine and bromine. Pools using ‘ozone with chlorine’
treatment systems, or other combined treatments, requirea special testing
procedure. See separate leaflet.