■ The warning symbols and their glossary are intended to ensure 

safe and proper use of the device by the user and prevent harm to 
the user and others.

■The meaning of the warning symbols are as follows:

b)Pulse width: 
TENS: 120µs
c)Impulse waveform: square wave
d)The product output has no the DC component.
e)Influence of output end open circuit and short circuit: it is able to 
support the influence of output end open and short circuit, and its 
performance will not be allowed to weaken.
f)Adjustment of output amplitude: 0-16 levels.
g)The timer of therapeutic equipment: 15 minutes.
h)Safety classification: internal electric source class and type BF 
equipment
i)Boundary dimension: 

Main unit : Model KTR-2492: Φ49.8*12.48

j)Product software version No.: A/0
k)Service life: the shelf life of subject device is 2 years; the use life 
of the Electrode Pad is 80 times and shelf life is 2 years 

Product Environmental Requirement

a)Normal work environmental requirements:
Environment temperature: +5

-+40

Environment humidity: 0%-80%RH; 
Atmospheric environment conditions: 860hPa-1060hPa.
b)Storage environment requirements:
Environment temperature: -25

-+55

Environment humidity: 0-93%RH; 
Atmospheric environment conditions: 500hPa-1060hPa.
c)Transport environment requirements:
Environment temperature: -25

-+55

Environment humidity: 0-93%RH; 
Atmospheric environment conditions: 500hPa-1060hPa.

Indicates potential danger, injury or serious injury due to 
incorrect usage

Indicates possibility of personal injury or damaged to goods 
through incorrect usage.

--  To avoid burns or damaged apparatus The device must not be used 

with Heart Fibrillation (HF) apparatus.

-- If the patient uses the therapeutic apparatus and HF apparatus at the 

same time, the contact plate may cause burns, it may also damage the 

apparatus. If the apparatus is used near (within 1 metre) of short-wave or 

microwave therapeutic apparatus, the output of the device may become 

unstable.

-- There is an increased risk of heart fibrillation if the electrode pads are 

used close to the chest.

-- Do not modify this equipment without authorization of the manufacturer 

--Do not attempt to replace the lithium batteries. Contact aftersales staff 

designated and authorized by the manufacturer directly.

-- Battery replacement should only be attempted by professional 

maintenance staff.

--The must not be used by people who can no longer take care of 

themselves, with no ability to express their own consciousness, or 

children.

Warning

attention

attention

Warning Sigh

Meaning

Indicates potential danger, injury or serious injury due to 
incorrect usage 

               Usage 
contraindication

13

12

8. Safety Precautions

Содержание KTR-2492

Страница 1: ...uk Tel 0800 953 1666 E mail enquiries tower health co uk Transcutaneous Electronic Nerve Stimulator 20 12 09 L180 x W125mm 128G CE KTR 2492 24 4 p a in g o n e Aegis model KTR 2492 Aegis F a s t e f f...

Страница 2: ...n obtained 3 Those who are or suspect they may be pregnant 4 Those with haemorrhage after acute trauma or fractures or those recovering from surgery 5 Those with skin particularly sensitive to heat or...

Страница 3: ...atural pain relief mechanisms The device additionally delivers pleasant massage like stimulation Aegis can be worn discreetly under clothing and activated whenever needed via the dedicated remote cont...

Страница 4: ...stic cover from side A of the gel pad Apply the gel pad to the Aegis unit as shown in the diagram Now remove side B from the gel pad Retain this plastic cover as it will be re used after treatment it...

Страница 5: ...ly to cycle through different combinations of pulses to find a mode that you prefer There are 6 modes in Automatic function Each time you switch Function or mode the device intensity will reset to lev...

Страница 6: ...h brand new gel pads Only use official Aegis gel pads with the Aegis device Only use one pair of pads per user of the device To purchase replacement gel pads contact Tower Health Ltd www tower health...

Страница 7: ...ged The speaker is damaged The battery is running low or running out The gel pad is damaged or dirty has no stickiness or is making poor contact with the skin Skin is too oily Recharge the battery Wip...

Страница 8: ...a 1060hPa c Transport environment requirements Environment temperature 25 55 Environment humidity 0 93 RH Atmospheric environment conditions 500hPa 1060hPa Indicates potential danger injury or serious...

Страница 9: ...s such as electrocardiograph Do not use the product in strong direct sunlight or where there is high heat inflammable material electromagnetic radiation and or high humidity Do not disassemble repair...

Страница 10: ...maging in operating rooms near active AF surgical equipment electrophysiology laboratories armored rooms or areas where short wave therapy equipment is used Do not use the system around strong electri...

Страница 11: ...C 61000 4 4 3V 0 15 MHz 80 MHz 6 Vm in ISM bands between 0 15 MHz and 80 MHz 80 AM at 1KHz 3V 0 15 MHz 80 MHz 6 Vm in ISM bands between 0 15 MHz and 80 MHz 80 AM at 1KHz Conducted disturbances induced...

Страница 12: ...217 Hz 2 2 0 3 0 3 0 2 0 3 0 3 28 28 9 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE 1 3 4 25 Band UMTS Pulse modulation 217 Hz 2 28 2450 2400 2570 Bluetooth WLAN 802 1 1 b g n RFID 2...

Отзывы: