Guidance and manufacturer’s declaration – electromagnetic immunity

The models KTR-230 are intended for use in the electromagnetic 
environment specified below. The customer or the user of the models
 KTR-230 should assure that it is used in such an environment.

Immunity 
test

IEC 60601 
test level

Compliance
level

Electromagnetic 
environment – guidance

Voltage 
dips, short 
interruptions 
and voltage 
variations
on power 
supply
input lines
IEC 
61000-4-11

Power 
frequency
(50/60 Hz)
magnetic 
field
IEC 
61000-4-8

30 A/m

 30 A/m

<5 % UT
(>95% dip 
in U

T

.)

for 0.5 cycle 
40 % U

T

(60% dip 
in U

T

) for 5 

cycles 70% U

T

(30% dip in U

T

)

for 25 cycles 
<5% U

T

(>95 % dip in 
U

T

) for 5 sec 

Mains power quality 
should be that of a typical 
commercial or hospital 
environment. If the user of 
the models KTR-230
requires continued 
operation during power 
mains interruptions, it is
recommended that the 
models KTR-230 be
powered from an 
uninterruptible power 
supply or a battery.

Power frequency magnetic 
fields should be at levels 
characteristic of a typical 
location in a typical 
commercial or hospital 
environment.

Not 
applicable

NOTE    U

T

 is the a.c. mains voltage prior to application of the test level

Conducted 
RF
IEC 
61000-4-6

Radiated 
RF
IEC 
61000-4-3

specifications 
for 
ENCLOSURE 
PORT 
IMMUNITY 
to RF 
wireless 
communication 
equipment 
(Refer to 
table 9 of 
IEC 
60601-1-2:
2014)

3 Vrms
150 kHz to 
80 MHz

6 Vrms in 
ISM  
bands

10 V/m

80 MHz to 
2.7 GHz

385MHz-
5785MHz 
Test

 

specifications 
for 
ENCLOSURE 
PORT 
IMMUNITY 
to RF 
wireless 
communication 
equipment 
(Refer to 
table 9 of 
IEC 
60601-1-2:
2014)

Not 
applicable

Not 
applicable

10 V/m

80 MHz 
to 2.7 GHz

385MHz-
5785MHz 
Test 

Portable and mobile RF 
communications equipment
 should be used no closer 
to any part of the models 
KTR-230, including cables, 
than the recommended 
separation distance 
calculated from the 
equation applicable to
the frequency of the 
transmitter.

Recommended 

separation distance

d=[3,5/V

1

]×P

1/2

d=1.2×P

1/2

 80 MHz 

to 800 MHz
d=2.3×P

1/2

 800 MHz 

to 2.7 GHz

where P is the maximum 
output power rating of the 
transmitter In watts (W) 
according to the transmitter 
manufacturer and d Is the 
recommended separation 
distance in meters (m). 
Field strengths from fixed 
RF transmitters, as 
determined by an 
electromagnetic site survey,
a should be less than the 
compliance level in each 
frequency range.

b

Interference may occur In 
the vicinity of equipment 
marked with the following 
symbol:

Содержание Easy

Страница 1: ...r choosing Paingone Easy Please read this user manual carefully before use Model KTR 230 Distributor Medi Direct International Ltd Unit 24 Wilford Industrial Business Park Ruddington Lane Nottingham N...

Страница 2: ...should not use electrical stimulation unless specialist medical authorisation has first been obtained Thank you for choosing PainGone Easy This simple device can help you to manage your pain to maint...

Страница 3: ...Y USE OFFICIAL PAINGONE EASY ELECTRODE PADS WITH THIS DEVICE Please do not try to attach any other electrode pad to this product ONLY USE ONE PAIR OF PADS PER USER OF THE DEVICE To purchase replacemen...

Страница 4: ...y The device is damaged Replace the battery Replace with a new electrode pad Wipe skin with a wet towel or alcohol Allow to dry fully before using the device Contact the point of purchase The device d...

Страница 5: ...ours KTR 230 119A V1 0 4 Safety precautions The purpose of the warning signs and symbols in this user manual is to make sure that the user can use this product safely and correctly and thus prevent da...

Страница 6: ...bnormal blood pressure malignant tumor serious cerebrovascular disease acute disease or currently under doctor s treatment should consult a doctor before use this stimulator 2 Do not use this stimulat...

Страница 7: ...to cause any interference in nearby electronic equipment The models KTR 230 are suitable for use in all establishments including domestic establishments and those directly connected to the public low...

Страница 8: ...s the a c mains voltage prior to application of the test level Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 specifications for ENCLOSURE PORT IMMUNITY to RF wireless communication equipment Re...

Страница 9: ...rding to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150kHz to 80MHz d 1 2 P1 2 80M...

Отзывы: