666036-0000 V1.0
4. Operation
NOTE
- Before initial operation, the nebulizer cup
assembly should be cleaned following instructions
described in the “Cleaning” section.
WARNING
- Before connecting the power cord, make
sure the I/O (ON/OFF) switch is in the O (OFF) position.
The plug is also served to disconnect the device.
L’appareil est également muni d’une fiche de
connexion à l’électricité
.
4.1 Daily Use Operation
CAUTION
- The Pacifica Compressor is designed for
intermittent use only. Do not operate it continuously
for more than 30 minutes for a single use without
turning it off and following a cooling period for least 30
minutes.
1. Before each use inspect the Pacifica Compressor
and nebulizer cup assembly for damage or wear,
replace as needed.
2. Place the Pacifica Compressor on a table or other flat
stable surface. Be sure you can easily reach the
controls when seated. Do not use this device on the
floor.
3. With the power switch in the O (OFF) position, plug
the power cord into an appropriate electrical wall
outlet.
4. Connect one end of the tubing to the compressor
air-outlet connector.
5. Assemble the nebulizer cup and add indicated
medication to the nebulizer’s cup before use.
6. Attach the other end of the tubing into the air-inlet
connector founded at the bottom of the nebulizer cup.
7. Turn on the Pacifica Compressor by pressing the
power switch to the I (ON) position and begin
treatment.
8. If treatment needs to be interrupted, simply press
power switch to O (OFF) position.
9. When the treatment is complete, turn off the
compressor by pressing the power switch to O (OFF)
position and unplug the unit from the electrical outlet.
Equipment not suitable for use in the presence of a
flammable anesthetic mixture with air or with oxygen
or nitrous oxide. This device has no AP or APG
protection.
NOTE : L’équipement ne peut être utilisé s’il y a
risque de mélange d’un anesthésique inflammable
avec l’air ou l’oxygène ou oxyde nitreux.
4.2 Safety Overload
NOTE
- Do not exceed 30 minutes of continuous
operation.
1. The motor of this device has a built-in thermal
overload protector. Should the motor overheat, the
protector will automatically shut off the motor. Should
this occur, turn the I/O (ON/OFF) switch to the O
(OFF) position and allow the motor to cool down for
approximately 30 minutes before turn it on again.
2. If the overload protector shuts off the motor frequently,
you may have an unstable voltage situation.
3. If the unit shuts down and cannot restart, it may need
to be replaced. Call your provider immediately.
5. Cleaning
5.1 Compressor Outer Case Cleaning
WARNING
- Electric shock hazard. Do not remove
outer case of this unit. All disassembly and
maintenance of this unit must be done by a qualified
service technician. Refer servicing to qualified service
personnel.
WARNING
- This unit does not require oil. Do not
attempt to lubricate any internal parts.
WARNING
- Unplug unit before cleaning. Do not
submerge in water for cleaning.
1. Wipe the main unit with a damp cloth every few days
to keep it dust-free.
2. Do not use any powdered type cleaners or soap. Do
not submerge the unit into water.
5.2. Nebulizer Cup Cleaning
All parts of the disposable nebulizer, except tubing, should
be thoroughly cleaned after each use. If your doctor
specifies a specific cleaning procedure, follow that
procedure.
WARNING
-To prevent possible risk of infection from
contaminated medication, cleaning of the nebulizer is
recommended after each treatment.
1. After each use, disassemble the nebulizer cup, tee
and mouthpiece completely, wash all parts in warm
soapy water and rinse thoroughly. Allow to air dry.
Store in a clean dry location.
2. Once daily, after washing as above, soak nebulizer
components in a fresh solution consisting of one part
warm water and one part white vinegar for 30 minutes
followed by a warm water rinse.
NOTE
- The disposable nebulizer kit is for single
patient use only.
NOTE
– Please follow national requirements to
dispose the unit properly.
6. Storage
Keep the unit and accessories dry. Avoid direct sunshine.
7. Maintenance
7.1 General Information
1. This unit is oil-less. Do Not Lubricate.
2. Risk of electric shock. Do not disassemble the main
unit.
3. All maintenance must be performed by a qualified
service personnel.
7.2 Filter Change
NOTE
- Change the filter once every
month or when filter turns gray.
Please follow the below
instructions as right figures.
1. Open the filter cap.
2. Remove dirty filter with a small,
pointed object. Discard the filter.
3. Replace with a clean filter.
Additional filters should be
purchased from your provider.
4. Put the filter cap back.
CAUTION
-
Do not use cotton or
any other material as a filter. Do not wash or reuse the
filter. Do not operate the unit without a filter.
8. Troubleshooting
If your Pacifica Compressor Nebulizer fails to function,
consult the Troubleshooting guide below. If the problem
persists, consult your equipment provider.
Problem
Cause and Solution
Device
doesn’t
operate.
‧
Check if plug is properly fit into an
appropriate electrical outlet.
‧
When device has been run continuously
for over 30 minutes right before using, an
auto shut down may activate by built-in
thermal protector, cool down device for
30 minutes before next usage.
Weak
Nebulization
‧
Check for proper electrical voltage.
‧
Check tubing for blockage or air leakage at
connection to Pacifica compressor or
nebulizer cup, replace as needed.
‧
Check the nebulizer cup if it is properly
assembled and not damaged. If there is
any damage, replace as needed.
‧
Check if filter is too dirty, replace as
needed.
9. Specifications
Electrical Rating
(Note: Refer to the rating
label on the product)
100VAC,60Hz,
1.2A
120VAC,60Hz,
1.2A
230VAC,60Hz,
0.6A
100VAC,50Hz,
1.2A
230VAC,50Hz,
0.6A
230V-240VAC,
50Hz,
0.6A
Maximum Compressor
Pressure
≧
30 psi
≧
30 psi
Operating Flow Rate
≧
5.5 lpm
≧
5.0 lpm
Nebulization Rate
0.2 - 0.3 ml/min
Particle size
0.5 ~ 5 microns*
Classification
Class II.
BF equipment.
IPX0
No AP/APG protection.
Applied part
Mouthpiece or Nasal mask
Dimensions (W x D x H)
14.0
×
18.8
×
10.0 cm /
5.5”
×
7.4”
×
3.9”
Weight
1.8 kg / 4.0 lb
Fuse (see note 3)
T5A 250VAC for 120V system,
T1.6A 250VAC for 230V system
Operation: 10
℃
to 40
℃
/ 50
℉
to 104
℉
Storage: -15
℃
to 50
℃
/ 5
℉
to 122
℉
Temperature
Transport: -15
℃
to 70
℃
/ 5
℉
to 158
℉
Operation: 10% to 90%RH non-condensing
Storage: 10% to 90%RH non-condensing
Environment
Humidity
Transport: 10% to 90% RH non-condensing
Note:
1. * Values are determined by supplier’s nebulizer.
2. Consult distributor(s) for additional technical
documents.
3. The Nebulizer is fitted with a fuse as specified in
the Specification column, it is designed to be
maintained by the manufacturer or its agents.
Users are not allowed to open the cover to
replace a blown fuse.
MANUFACTURED FOR:
Drive Medical Design & Manufacturing
99 Seaview Boulevard
Port Washington, NY 11050