AFX SERIES® OPERATION MANUAL
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 141 of 485
6.6.6
Multiple User Waveforms in LIST Transients
Transient programs are very useful to deliver precisely controlled transitions between
different waveforms to a unit under test. This means transitions from a normal AC sine wave
to a non-sinusoidal or distorted waveform can be accomplished by using different waveform
at different segment or step entries.
Waveforms are numbered from 1 through 200 with 1 being a fixed sine wave. All other
waveform registers are user defined arbitrary waveforms.
When in AC mode, transient segments or steps can called out different waveform numbers
on each of up to three phases
. For each segment, up to six different waveforms can be
selected. The same waveform can be repeated as of often as needed within the same
transient program.
In three or split phase mode, each phase in a transient program can use its own set of up to
six user-defined waveform. Thus, up to 18 different waveforms are available when in three-
phase mode.
Figure 6-18: Available User Waveforms in Transients
5
Requires Front Panel Firmware revision 1.3.3.
Содержание AFX Series
Страница 484: ......