Query Format
PROGram[:SELected]:CHECK? [,<nr1>] [,<cr>]
Description
This command performs a check on the selected program steady
state tables looking for the following conditions:
•
Hardware Limit violations
•
User Limit violations
•
Saturation limits.
If no violations are present, this command returns OK and the
program can be executed without generating any errors. If the
selected program cannot be run due to one or more issues with its
content, one or more errors detailing the issues will be returned.
For example, if the upper voltage user limit is set to 100V and the
program contains a setting of 300V, a “Cannot load program, Run
PROGRAM:CHECK?” error will be generated when attempting to
execute this program. When sending the PROGRAM:CHECK? Query,
the response will be “VOLT1 AC voltage set point cannot change due
to user limit”.
Parameters
Option: PROGRAM NUMBER [,<nr1>]
Option: MEMORY [,<cr>]– Available are: INTERNAL | RAM | USB
Returned Data Format
<cr>
Query Example
PROG:CHECK?
VOLT1 AC voltage set point cannot change due to user limit
Query Format
PROGram:EXECuted:CHECK? [,<nr1>] [,<cr>]
Description
This command serves the same purpose as the PROG:CHECK?
Command but applies to a program that is executing.
Parameters
Option: PROGRAM NUMBER [,<nr1>]
Option: MEMORY [,<cr>]– Available are: INTERNAL | RAM | USB
Returned Data Format
<cr>
Query Example
PROG:EXEC:CHECK?
VOLT1 AC voltage set point cannot change due to user limit
Command Syntax
PROGram[:SELected]:DELete
Description
This command deletes the selected program. Attempting to DELete
an EXECuting Program will result in an Error.
Parameters
Program number
Parameter Format
<nr1>
Example
PROG:DEL 9