INTRODUCTION 

 
  

Welcome to SMART-E  programming system with SPL software, the application 
designed by Pacetronix to consult and modify the parameters that rule pacemaker 
activity belonging to their SSI / SSIR / DDD / DDDR / Bi-Ventricular pacemakers. 
 
SPL software, created to be run in a personal computer with Windows XP or Windows 
98 operating system, is connected by means of a programming interface to a small 
SMART-E Programming Head Model SEH-11, which communicates by telemetry to the 
pacemaker implanted in the patient. 
 
The  SMART-E  Programming Head Model SEH-11  must be placed over  the implant 
zone, making sure that the light blinks following its frequency.  The  SMART-E 
Programming Head Model SEH-11  strength is such that it can communicate with 
pacemaker when it is kept within  air gap of 50 mm height.  The button located in the 
SMART-E  Programming Head Model SEH-11  is usually used to "interrogate" the 
pacemaker, that is, to show in the SPL software screen the programmed values that 
determine it’s functioning. 
 
Using  SMART-E  Programmer Model SEP-11  with SPL software the Physician may 
interact with the pacemaker: modify the frequency and duration of the generated 
stimulus, measure pacing and sensitivity thresholds, study statistics and graphics of the 
assistance given by the heart stimulator, keep an electronic file of patients and their 
controls, calculate which is the battery consumption of each program, register standard 
programs for further use, reprogram the pacemaker in order to quickly control heart 
rhythm problems, etc. 
 
SPL software offers in a graphic interface of quick and easy handling, all these multiple 
options which improve the functioning of the pacemakers built by Pacetronix and 
achieve, in every case, accurate and personalized clinical follow-up. 

 

 

4.

 

TECHNICAL SERVICE 

 

ln case of additional help about SMART-E Programmer Model SEP-11, SPL software or 
about  SSI  /  SSIR  /  DDD  /  DDDR  /  Bi-Ventricular  pacemaker series you may consult 
Pacetronix through: 
 
Mail: Plot No.15, Sector-II, Pithampur, Dist. Dhar (M.P.) 454775, INDIA 
 
Telephone: 07292-411105 
 
Fax: 07292-2762728 
 
E-mail: 

pacetronix@

hotmail.com  

 
Web:  

www.pacetronix.com

  

Содержание SMART-E PROGRAMMER

Страница 1: ...rface Model SEP 11 Programming Head Model SEH 11 Shree Pacetronix Ltd Plot No 15 Sector II Industrial Area Pithampur Dist Dhar 454775 M P INDIA Telephone 07292 411105 Fax 07292 2762728 E mail pacetron...

Страница 2: ...Revision Date Change Description 01 January 15th 2017 IFU 02 August 25th 2020 User Information...

Страница 3: ...planation 1 Operating instructions 2 Do not dispose of this product in the unsorted municipal waste stream Dispose of this product according to local regulations 3 Product documentation 4 DEFIBRILLATI...

Страница 4: ...WARE SCREEN 13 8 1 Use of the Mouse and Keyboard 13 8 2 Run SPL software 14 8 3 SPL software Screen 14 8 4 Actions 17 9 DATA LOG BROWSER 23 9 1 Implant 23 9 2 ECG 24 9 3 Reports 24 10 Tables 25 A Para...

Страница 5: ...11 CLEANING 31 12 SAFETY TECHNICAL CHECKS 32...

Страница 6: ...ho do not have previous experience in the use of a personal computer that runs Windows XP or Windows 98 operating system It defines the terms used throughout the manual and explains the usual working...

Страница 7: ......

Страница 8: ...the programmed values that determine its functioning Using SMART E Programmer Model SEP 11 the Physician may interact with the pacemaker modify the frequency and duration of the generated stimulus me...

Страница 9: ...Shree Pacetronix Ltd shall not be liable for any medical expenses adverse body reactions or medical complications or other direct or consequential damages resulting from the use of SMART E Programmer...

Страница 10: ...rogrammed values that determine it s functioning Using SMART E Programmer Model SEP 11 with SPL software the Physician may interact with the pacemaker modify the frequency and duration of the generate...

Страница 11: ...he connection of USB cable from both PC and Interface end ii SMART E Programming Head Model SEH 11 connection Connect SMART E Programming Head Model SEH 11 male connector to programming interface at F...

Страница 12: ...PC a green LED on LED bar glows and indicating power up of SMART E Programmer Model SEP 11 II Signal Strength Indicator LED bar placed on the SMART E Programming Head Model SEH 11 indicates the telem...

Страница 13: ...ents indicated above and insert the CD labeled SPL software into your CD ROM drive Double click the setup file in the folder Once the Auto Run Installation program begins just follow the instructions...

Страница 14: ...WARE SCREEN This section describes the general characteristics of the SPL software Screen and Windows XP 98 the built in operating system as well as some conventions applied throughout this manual If...

Страница 15: ...d it means that the application has been installed within a program group then all the folders of the option PROGRAMS have to be searched using the mouse until SPL software is found in one of them The...

Страница 16: ...uld be verified that the lighin the SMART E Programming Head Model SEH 11 is blinking when it is located in the implant zone indicating that the distance to the pacemaker is correct Program button all...

Страница 17: ...arrows In the right part of the Status bar there are three boxes that indicate if the lock keys in the keyboard have been activated CAPS The first box indicates if the key Caps Lock has been activate...

Страница 18: ...options Finally the dialog boxes as the windows have a title bar from which they can be dragged throughout the screen However they have as a difference a fixed size that cannot be changed 8 4Actions...

Страница 19: ...etry by clicking on the Retry button of the message box If the interrogation is successful the different parameters of the pacemaker which vary according to the model appear on the screen programming...

Страница 20: ...ROGRAMMING option If the programming action was performed successfully the message Programming OK will appear otherwise an error message box will be shown In this case in order to repeat the action yo...

Страница 21: ...shold a rate that allows the patient to dominate the rhythm must be selected START button must be clicked to start the threshold detection to increase the Sensitivity parameter NEXT button should be c...

Страница 22: ...ard the procedure is the same as the one described Longitivity Estimated service life Dialog box for estimate the service life Using the Longitivity dialog box the user can vary the rate mode pulse wi...

Страница 23: ...h the total number of events pacing and sensing The Graphic button shows the different values in a bar graphic format The Reset button zeroes all the statistical values in the pacemaker and the Close...

Страница 24: ...registers shown in the Statistics command SPL software stores in a data base the values used to program the pacemaker in each control or visit as well as personal and specific implant data of the pati...

Страница 25: ...arker and ECG data ECG Recording ECG of a patient can record by clicking ECG RECORDING Button ECG Gain 3 gains 0 1 2 can be set using the slider Right Corner ECG Sensitivity ECG sensitivity can be adj...

Страница 26: ...onfiguration for sensing Unipolar Bipolar Sensing Hysteresis 0 to 20 in steps of 5 Hysteresis Refractory Periods From 200 to 500 ms in steps of 15 625 ms Refractory Upper Rate From 80 to 180 ppm Trigg...

Страница 27: ...ge When the rate is greater than 120 ppm the refractory period decreases automatically to 250 ms ARC C Parameter Values of DDDR 747R Pacemakers Parameter Valid values Title on the screen Mode DDD VDD...

Страница 28: ...From 1 to 15 Slope Reaction time From 10 to 60 Sec in steps of 10 Sec Up Time Recovery time From 1 to 10 Min steps of 1 Min Dn Time Automatic Refractory Change On Off When programmed On if the rate is...

Страница 29: ...lse Amplitudes A RV Pulse Amplitudes LV 2 5V 5 0V 5 0 V Pacing Electrical configuration for pacing Unipolar Bipolar Pacing Sensitivities For Atrium From 0 5mV to 4 0 mV For Right Ventricle From 1 0 to...

Страница 30: ...harak DR Trinity 3000 Mode Yes Yes Rate Yes Yes Upper rate Yes No Backup mode No No Blanking Yes Yes Pacing Yes Yes Sensing Yes Yes Refractory Yes Yes PVARP Yes Yes Coupled No No AV Delay Yes Yes AV D...

Страница 31: ...n 140 ppm S 0 80 Pacemaker sensed events amount at a rate in the range 0 and 80 ppm S 80 100 Pacemaker sensed events amount at a rate in the range 80 and 100 ppm S 100 120 Pacemaker sensed events amou...

Страница 32: ...rate in the range of 120 and 140 ppm S 140 Sensed events percentage related to the sum of paced and sensed events that occurred at a rate in the range higher than 140 ppm Others Times Noise Number of...

Страница 33: ...l to remove all visible dust marks The cables may be immersed for cleaning The cables must be thoroughly dried after cleaning Inspection and testing by a qualified technician should be done after clea...