1450 First Ave.
Page 3 of 10
Chippewa Falls, WI 54729
(800) 283-1141 or (715) 723-1141
Fax: (715) 723-7890
DA40T User Manual
Pacer Instruments, Inc
Rev 1.0
www.pacer-instruments.com
11-Aug-2007
Phone: 1-800-283-1141
INTRODUCTION
Pacer’s model DA40T digital thermometer-anemometer is a versatile instrument for measuring air velocity and
temperature of airflow from HVAC ducts or process air flow.
The heavy, all metal (except electronics) probe can be used for airstreams which have a wide range of humidity,
temperature and contaminants without compromising accuracy. Temperature sensor is a Platinum Resistance
Element (RTD).
Features include choice of probe diameters, custom cable lengths, tolerance of temperatures up to 210
˚
F
(98.9
˚
C) at the probe, and durability. Optional special-purpose temperature-only probes may be purchased from
factory.
SECTION 1 - SPECIFICATIONS
Range:
APT275 probe:
40 – 7800 FPM (0.2 - 40.0 MPS)
APT100 probe:
60 – 6800 FPM (0.3 – 35.0 MPS)
Temperature – Either velocity probe:
-22
º
to 212
º
F (-30
º
to 100
º
C)
Temperature – Optional temperature-only probes:
-139º to 392.0
º
F (-95º to 200.0
º
C)
Accuracy:
Air Velocity:
±1.0% of reading ±1 digit
Temperature:
±0.3% of Reading ±1 digit
Resolution:
Air Velocity:
1 FPM or 0.01 MPS
Temperature:
0.1
º
F or
º
C (1
º
F below -99.9
º
F)
Operating Temperature:
Instrument:
32 to 125
º
F (0 to 50
º
C)
Probes:
-4
º
to 210
º
F (-20
º
to 98.9
º
C)
Power Supply:
2 AA alkaline batteries, E91 Eveready or equivalent
Battery Life:
Approximately 200 hours
Battery check:
Automatic low battery display
Display:
0.5” LCD, 4 digits
Weight:
8 ounces with batteries
Dimensions:
Instrument:
7.1” x 3.0” x 0.8”
APT275 probe:
2 ¾” diameter
APT100 probe:
1” diameter