4. 

Depress the foot pedal again and now look at the control panel. 
The red 

“x-ray on” light should illuminate and remain on for as long 

as you hold down the foot switch.  Release the foot pedal. 

 

5. 

Place a component (such as a circuit board) inside the inspection 
chamber directly underneath the x-ray tube. 

 

6. 

Depress and hold the foot switch.  The x-ray image of the 
component should now be visible. 

 

7. 

Move the component while the foot switch is depressed. If the 
component is large enough to extend beyond the curtains, it is 
permissible to move the component while the foot switch is 
depressed- AS LONG AS THE CURTAINS DO NOT OPEN 
DURING MOTION, AS THIS WILL INCREASE X-RAY LEAKAGE. 
The image should move with the component.  The image will 
display a trail behind the component, which will disappear when the 
component has stopped moving.  This is a normal occurrence and 
is a function of the image processor.  The video processor included 
with this system is preset to 8 frames averaging.  Averaging makes 
the x-ray image less grainy and can be adjusted to improve image 
quality but with an increase in motion lag. 

 

8. 

Images may be captured electronically when the XR 3500/3700 is 
used with the TF 1700/2700 PC or IR 3000 PC and software. 

 

*Your system is now ready for use*

 

 

9.  XR 3500/3700 Maintenance

 

 

The XR 3500/3700 system has been designed for simple maintenance. The only 
suggested maintenance is to keep the unit visibly clean and to keep the system

’s 

calibration up to date. The window is made of a lead-impregnated acrylic resin; 
NEVER allow solvents that will attack acrylic to be used near the window. If you 
find that your system is coming close to its calibration date, please contact 

PACE 

to schedule a calibration. 

 

 

 

It is very important that you contact 

PACE 

early, so that we may schedule 

service at a convenient time. 

 

 

Warning: 

The acrylic and vinyl around the opening of the XR 3700 

contain lead. Always wash your hands immediately after use. Eating and 
drinking without washing your hands may cause you to ingest lead. *

 

Содержание XR 3700

Страница 1: ...Ray Inspection System PN 8007 0542 120V PN 8007 0546 230V XR 3500 Real Time X Ray Inspection System PN 8007 0540 120V PN 8007 0544 230V This manual operation is for PACE s XR 3500 50kv unit and PACE...

Страница 2: ...y Guidelines 6 XR 3700 Specifications 7 Features 7 System Operation 8 Set up 9 XR 3700 Initial Power up and Operational Test 10 XR 3700 Maintenance 11 XR 3700 Troubleshooting 12 Design Safety Standard...

Страница 3: ...aware that the ionizing radiation x rays can constitute a distinct hazard if not employed in strict accordance with instructions provided in this manual for maximum operator safety Exposure to excessi...

Страница 4: ...e a full understanding of how it works Additionally they should be familiar with established radiation safety exposure practices sanctioned by the National Bureau of Standards Handbook X ray Protectio...

Страница 5: ...d constraints regarding the modification of x ray systems conforming to 21CFR mentioned above Typically an x ray system must be registered with the appropriate state agency by the physical possessor o...

Страница 6: ...ount of radiation being emitted by the machine is less than 0 5mR hour We recommend performing a radiation survey Following initial installation before placing the x ray system into service Whenever t...

Страница 7: ...PCB size 760mm x unlimited mm 30 x unlimited Focal spot 0 2mm 0 008 Focal spot to image plane distance 124mm 4 875 Adjustments Live or Capture video signal Frame averaging up to 32000 frames Video Ga...

Страница 8: ...etermine the presence and degree of interlayer shifts At the onset of either conventional or small hole drilling inspection is used to qualify hole to pad alignment In addition to enhancing product qu...

Страница 9: ...e bnc cable should be connected to the composite video input on the video monitor a bnc to RCA converter might be needed The power cord should be connected to a properly grounded 120 VAC power supply...

Страница 10: ...as been returned to the off position The power light should blinks when the key switch is on and the cover lock is disengaged Then when the cover lock is engaged the power light turns steady 2 Make su...

Страница 11: ...eo processor included with this system is preset to 8 frames averaging Averaging makes the x ray image less grainy and can be adjusted to improve image quality but with an increase in motion lag 8 Ima...

Страница 12: ...t the system Power The power category covers all problems that include lack of power to a component or to the full system Power problems would include The system not powering up The x ray controller s...

Страница 13: ...r Devices and Radiological Health Branch of the FDA as an Analytical X ray System As registered the system has a number of safety features provided to minimize any x ray scatter reaching the operator...

Страница 14: ...ecific regulations regarding the monitoring of possible radiation leakage of industrial x ray cabinets are determined by the individual state or country There are a number of measures which can be tak...

Страница 15: ...ption repair or replace any defective products at its facility or other location approved by it at no charge to user or provides parts without charge for installation by the user in the field at user...

Страница 16: ...orporated to obtain the latest specifications The following are trademarks and or service marks of PACE Incorporated MD USA INSTACAL FUMEFLO HEATWISE PACEWORLDWIDE PERMAGROUND POWERMODULE and TEMPWISE...

Отзывы: