©2022 PACE Inc., Vass, North Carolina, All Rights Reserved Page 8
https://paceworldwide.com/
9. The LED Display now reads "End". The Set-Up Mode procedure is now complete.
Choose one of the following steps:
•
Press and release the Scroll Up Key ( ) to save the changes, exit the Set-Up
Menu and return to normal operation.
•
Press and release the Scroll Down Key (
) to return to the start of the Set-Up
Menu without saving any changes. Go back to step 3, Password Entry.
Corrective Maintenance
Display Message Codes
Display
Message
Description
“no”
The incorrect password has been entered. The displayed message will time
out after 6 seconds and revert to normal operation. Enter the correct
password.
“OCE”
“Over Current Error” The Heater circuit is shorted, or the handpiece has
failed. Inspect handpiece and heater element, replace as needed.
“OSE”
“Open Sensor Error” Unit is unable to sense the sensor from the handpiece.
Make sure handpiece is connected. Inspect handpiece and heater element
for damage or wear, replace as needed.
“OFF”
The AutoOff feature has activated, turning off the handpieces. Press any
button on the unit to return to normal operation.
Power Source Issues
Most malfunctions are simple and easy to correct.
Symptom
Probable Cause
Solution
No power to
system
Blown Fuse
Replace the fuse (located in the AC
Receptacle Fuse Holder) with one of the
same.
Handpiece not
heating up
Defective Heater Element
Refer to the appropriate handpiece
manual.
Power Source Malfunction Contact PACE
Spare Parts
Description
PACE Part Number
Fuse 1 Amp, Time Lag (for 115V)
1159-0246-P5
Fuse 0.50 Amp, Time Lag (for 230V)
1159-0213-P5
Optional under shelf mounting bracket
1321-0609-P1
Service
Please contact PACE or your local distributor for service and repair.