Correctiv
e Maintenance
Power Source
Most malfunctions are simple and easy to clear. Refer to Table 4 below to clear these malfunctions. If you
encounter any difficulty clearing the malfunction, contact PACE Technical Support directly at Tel. (toll free)
1-888-535-PACE(7223)or(301)490-9860,FAX(301)483-7030.
Symptom
Probable Cause
Solution
Blown Fuse (F1)
Replace Fuse F1 located on rear of Power
Source in the AC Receptacle.
Digital Readout is blank.
No LEDs on. No motor.
Defective handpiece.
Disconnect all handpieces. Check each
handpiece using the applicable handpiece
manual or Table 5 of this publication.
No handpiece connected to Plug handpiece into CH 1, CH 2 OR CH 3.
power source.
E-1 displayed on Digital
Readout.
Open sensor in handpiece.
Refer to handpiece Operation Manual for
Corrective Maintenance procedures.
E-2, E-3,E-4 or room
Defective handpiece.
Disconnect the handpiece connected to
temperature displayed on
the current channel (channel displayed on
Digital Readout.
Digital Readout). Check the handpiece
using the applicable handpiece manual or
Table 5 of this publication.
Insufficient AUTO
Filter(s) and/or handpiece(s) Refer to applicable handpiece manual(s) for
SNAP-VAC (vacuum) or air
require corrective
instructions on performing proper
pressure.
maintenance.
"Corrective Maintenance" procedures.
Table 4. Corrective Maintenance, Power Source
52