SET-UP
NOTE
If more than one air-operated handpiece is connected
to the power source, insure that only one of the Air
Hoses is connected to either the VisiFilter assembly
(connected to the AUTO SNAP-VAC Port) or the
Controllable PRESSURE Port. Attachment to both
simultaneously will cause a deterioration in
performance.
TRADITIONAL METHOD (CONT'D)
5.
Connect the handpiece power cable connector plug to one of
the Power Output Receptacles. For convenience, PACE
recommends the use of CH 3 for air handpieces.
Figure 9. Handpiece Connection
QUICK CONNECT METHOD
May be used with any handpiece except the TJ-70
ThermoJet.
1.
Prepare a VisiFilter in the following manner.
a)
Connect a 1 inch (2.5cm) length of clear pvc Air Hose to
each side of the VisiFilter; push and turn hose onto
VisiFilter nipple to seat.
b)
To the free end of the Air Hose connected to the FLOW
IN side of the VisiFilter, insert the ribbed end of a Female
Quick Connect Hose Mount Fitting (P/N 1259-0086).
c)
Insert the ribbed end of a Male Quick Connect Hose
Mount Fitting (P/N 1259-0087) into the free end of the Air
Hose connected to the FLOW OUT side of the VisiFilter.
Figure 10. VisiFilter Preparation
22