I
NSTRUCTION
M
ANUAL
TERAPRESS TP 7500S
Mounting Press
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1
-
520
-
882
-
6598 Fax +1
-
520
-
882
-
6599 email: [email protected] Web: https://www.metallographic.com
Table 4
-
4 Conductive Mounting Powders
* Grams of powder to produce a
5/8”
mount without any sample
20
Parameters
1
-
inch
mounts
1.25
-
inch
mounts
1.5
-
inch
mounts
2
-
inch
mounts
Temperature (ºC)
160°C
(320°F)
160°C
(320°F)
160°C
(320°F)
160°C
(320°F)
Machine gage pressure (psi)
MPa
2150
14.8
2500
17.2
3750
25.9
6250
43.1
Time (minutes)
-
minimum
8
10
12
12
Grams of powder*
13.7
21.4
30.8
55
Table 4
-
5 Acrylic Mounting Powders
Parameters
1
-
inch
mounts
1.25
-
inch
mounts
1.5
-
inch
mounts
2
-
inch
mounts
Temperature (ºC)
160°C
(320°F)
160°C
(320°F)
160°C
(320°F)
160°C
(320°F)
Machine gage pressure (psi)
MPa
2150
14.8
2500
17.2
3750
25.9
6250
43.1
Time (minutes)
-
minimum
8
10
12
12
Grams of powder*
13.7
21.4
30.8
55