
27
3.
Wet the dry soft cloth with 75% alcohol, and then wipe all surfaces of the motor
component,
power adapter, charging base and other components for 3 minutes.
4.
Put the motor component, power adapter, charging base and other components
back
into
the clean storage area.
Note:
•
The cleaning and disinfection must be performed within 10 minutes before use.
•
The disinfectant used must be used immediately, no foaming is allowed.
•
In addition to 75% alcohol, you can use non-residue disinfectants such as Oxytech from
Germany, but you must respect the concentration, temperature and time specified by the
disinfectant manufacturer.
6.4 The cleaning, disinfection and sterilization of Outer Sheath
;
unless otherwise
stated,
they will be hereinafter referred to as “products
.
”
•
The use of strong detergent and disinfectant (alkaline pH>9 or acid pH <5) will reduce the life
span of products. And in such cases, the manufacturer takes no responsibility.
•
The products may not be exposed to temperature above 138
°C (280°F)
.
•
Processing limit
The products have been designed for a large number of sterilization cycles. The materials
used in manufacture were selected accordingly. However with every renewed preparation for