18
(so called R-curve) should be selected when
use of SpO
2
simulator, e.g. for Index 2 series
SpO
2
simulator
from
Fluke
Biomedical
Corporation,
please
set
"Make"
to
"DownLoadMake: KRK", then the user can use
this particular R-curve to test the Oximeter. If
the SpO
2
simulator does not contain "KRK"
R-curve, please ask the manufacturer for
helping to download the given R-curve into the
SpO
2
simulator.
High-pressure sterilization cannot be used on
the device.
Do not immerse the device in liquid.
It is recommended that the device should be
kept in a dry environment. Humidity may
reduce the life of the device, or even damage
it.
7.2 Cleaning and Disinfecting Instruction
Surface-clean sensor with a soft cloth damped
with a solution such as 75% isopropyl alcohol,
if low-level disinfection is required, use a mild
bleach solution.