1.  Cautions 

Please  read  these  Operating  Instructions  carefully.  This  device  is  only  to  be  used  by  fully  trained 
personnel. Ensure that the device is tested for  electrical safety at least annually. The pipeline system 
should be subjected to preventative maintenance to ensure the systems

’ reliability. 

To  operate  the  Automatic  Twin  Cylinder  Changeover  System  correctly,  ensure  that  the  Operating 
Instructions in this document are followed. Ensure the cleaning procedure and safety precautions are 
followed  to  avoid  the  risk  of  fire  and  explosion.  Contact  our  Service  Department  immediately  if  an 
electrical fault has been detected when using the Automatic Twin Cylinder Changeover System.  
Inspect the condition of the pipeline for any sign of damage i.e. probes, tubing, outlets crimp connections 
etc. 

 

Ensure that you check the contents of your package thoroughly before use. 

Ensure the cable is neatly wrapped prior to moving.   

 

Please  avoid  using  the  unit  near  other  types  of  electrical  equipment  that  are  non-approved  electrical 
medical devices that may release strong electromagnetic/radiated emissions. 

 

DO NOT

 stretch, pull or force the connecting hose through an acute bend. 

DO NOT

 use any flammable agents near or on the equipment when in use. 

DO NOT 

subject the system to excessive amounts of water or cleaning fluid or allow surfaces to become 

excessively wet. 

DO NOT

 jet wash the product or use in severe / wet environments. 

DO NOT

 clean the system whilst pressurised. 

DO NOT

 stretch  or  pull the  12  volt  DC  cable and  avoid  contact  with  any  sharp  implements  that  may 

penetrate the cable. 

DO NOT

 remove the front panel of the Automatic Twin Cylinder Changeover System for any reason. If 

the unit fails to operate you MUST contact the Oxylitre Service Department for service recommendations.  
The  panels  on  the  product  must  only  be  opened  by  an  Oxylitre  Service  Engineer  or  by  qualified 
personnel. 

2. 

Introduction 

The  unit  has  been  specifically  designed  for  use  with  a  twin  supply  pipeline  system.  The  Automatic 
Changeover System is available for both Oxygen, ENTONOX and Medical Air 4 Bar.  

This system manages the Change-over process from one cylinder to another automatically, switching 
over to the other cylinder when the cylinder pressure is low.  

The

 

Automatic  Changeover  Systems  robust  design  features  a  flame  retardant  case,  with  green  flow 

direction indication  LED’s,  red  empty cylinder  LED’s,  a  buzzer/alarm  to  indicate that  the system  has 
switched cylinders  or  no  gas  is  available. The system  is  also Ambulance  Control Module  compatible. 

Содержание AP4600 MK3 Series

Страница 1: ...Automatic Twin Cylinder Changeover System AP4600 Series MK3 Instruction for Use Made In the UK 2797 Oxylitre Holdings Limited ...

Страница 2: ... a cylinder Button 5 Operating the Unit 5 1 Electrical Connection 5 2 Start up Instruction after Power Down 5 3 Safety Precautions for the Prevention of Fire Explosion 5 4 Cylinder Connection 5 5 Audible Alarm Function and Visual References 5 6 Automatic Changeover 5 7 Over Pressure Safety Features 6 Servicing Preventative Maintenance Cleaning 6 1 Pipeline Inspection 6 2 Cleaning 6 3 Service Repai...

Страница 3: ... bend DO NOT use any flammable agents near or on the equipment when in use DO NOT subject the system to excessive amounts of water or cleaning fluid or allow surfaces to become excessively wet DO NOT jet wash the product or use in severe wet environments DO NOT clean the system whilst pressurised DO NOT stretch or pull the 12 volt DC cable and avoid contact with any sharp implements that may penet...

Страница 4: ...t 2 2 1 Pre Use Inspection Remove all packaging from your Automatic Changeover System inspect all parts for any signs of damage and ensure all of the contents are there If you see any signs or suspect that the unit may have been damaged DO NOT use Notify Oxylitre immediately 2 2 Contraindications None known 2 3 Warranty The Company Warranty in respect of durable goods but not in respect of disposa...

Страница 5: ...tre Twin Cylinder Changeover System User Interface Cylinder Selection Supply Graph Test Button Primary Supply Press to Select Cylinder Cylinder Status LED Alarm Mute Secondary Supply Outer Case System line Pressure Indicator ...

Страница 6: ...st mode 4 3 System Over Pressure Indicators If the active cylinder over pressurises the System Pressure Display will display pressure until it registers 5 9 bar then it will then flash HI on the display A repeating alarm will sound and the cylinder status LED will flash Red The unit will the automatically switch to the other cylinder the alarm will stop but the cylinder status LED for the over pre...

Страница 7: ...n allows the user to silence the alarm This is useful when removing cylinders carrying out maintenance or purging the system Pressing the Alarm Mute Button or the Test Button will automatically bring the audible alarm back online After maintenance work replacing an empty cylinder or when the system is fully operational this button is also used to allow the unit to reset the system The unit will pe...

Страница 8: ...mary cylinder on a regular basis Step 1 Connect the hose tail on the primary cylinder side to the primary cylinder Do not turn on the cylinder Step 2 Connect the hose tail on the reserve secondary cylinder side to the reserve cylinder Do not turn on cylinder Step 3 Turn on the primary side cylinder and pressurise the system The system will perform an input availability inspection test as mentioned...

Страница 9: ... In the event of this happening the Cylinder Status Indicator will be illuminate red and the Press to select a cylinder Button will be locked out on the failed cylinder side To bring the failed side back on line change the cylinder and reset the unit reference 4 6 5 6 Automatic Input pressure inspection If the vehicle system is to be left unattended overnight it is advised to turn off the cylinder...

Страница 10: ...lity 6 1 Pipeline Inspection once per annum minimum Inspect the condition of the Pipeline for any damage i e damage to the probe tubing ferrules outlets etc Also carry out routine testing and general safety protocols in accordance with HTM2022 6 2 Cleaning to be carried out on a regular basis The Pipeline system and Automatic Twin Cylinder Changeover unit requires cleaning on external surfaces onl...

Страница 11: ...on Mentally Stable Level of Competence Minimum Medically Competent No maximum Patient Population Age new born to adult Nationality Multiple Weight 2 5 Kg Application Environment Ambulance road vehicle Hospital Not suitable for use in the presence of flammable liquids or gases Not intended for outdoor use can withstand spraying of water IP53D Protected against spraying water from at up to 60 degree...

Страница 12: ...rotected against spraying water from at up to 60 degrees from vertical Protected against the access with a wire Performance Units of Measurement Bar Pressure Range 3 0 to 6 0 Accuracy 5 on indicated value Flowrate 40lpm minimum delivery Permitted Leakage Zero Unit Dimensions Height 210mm Width 145mm Depth 73mm Weight 1 5 Kg excluding Hose Tails Electronics Mains Input Voltage 12 17 VDC nominal 0 5...

Страница 13: ...9 Materials Specification Enclosure Front Fire Retardant ABS Rear Mounting Panel 6026 T9 Aluminium BS5682 Probe 303 Stainless Steel ACU Keypad Membrane Polyester PCB Mounting Bracket Aluminium PCB Mounts Female Female Delrin Main Manifold Assembly Block Brass Steel Nitrile Micro Valve Manifolds Aluminium Screws Steel Hose PVC 10 Disposal Not for general waste At the end of life cycle the Automatic...

Страница 14: ...he product has been designed to comply with the safety and design requirements of the European Directives for Medical Devices including Council Directive 93 42 EEC The product complies with the requirements of the relevant Safety European Automotive Directives for use in Ambulances and other Medical Vehicles Follow instructions for use Indicates Caution Waste Electrical Electronic Equipment WEEE c...

Страница 15: ...14 Manufacturer Date of Manufacture Fragile Keep dry Temperature limit Doc Ref Issue No Date Doc OP 2805a 1 0 DCN00911 25 05 2021 ...

Отзывы: