There are two columns (metal tubes) in the concentrator located 

behind the particle filter. These two columns will be removed and 

replaced when service is required. Ensure adequate time is allowed 

to complete all steps without disruption is alloted when performing 

this maintenance.

Lay the concentrator down on its side.

Press the two latch buttons on the sides of the device to 

remove the cover from the bottom of the concentrator 

in order to access the two columns.

Press the latch button with thumb or 

finger and pull the column out of the

concentrator using the pull ring attached 

to the bottom of the column cap.

Repeat steps 5-6 to remove the 

other column.

Remove the column completely 

from the concentrator.

Remove the upper and lower dust 

caps from each column. There 

should be a total of two dust caps 

removed from each column. Make 

sure there is no dust or debris 

where the dust caps were located.

Insert the new column into the concen-

trator. Do not leave the column exposed 

after the dust caps are removed; it should 

be inserted into the concentrator  

immediately to minimize  

environmental exposure.

The spring loaded latch button should fully 

return to the closed position when the 

column is completely inserted.

Continues on back

Installation of the new columns (metal tubes):

Replace the lower cover and return the concentrator 

to its upright position.

PMS 661 main color/651 swish

3

4

5

7

6

8

9

10

12

www.inogen.com

1

2

Unplug the concentrator from the electrical outlet.

Turn the concentrator off by pressing the power button  

to shut down the device.

 NOTE 

 

Column change instructions are only to be used when maintenance is 

   

required and are not intended for practice purpose. 

11

Repeat steps 9-10 to install the other column.