10. INTERCHANGEABLE ROTOR
ROTOR_8 SLOT
8 X 0.2,0.4,0.5, 1.5,2 ml
11. CHANGING THE ROTOR
PCRROTOR
16x 0.2 ml
Insert the hexagonal allen key into the marked hole provided on the side
of the rotor. The grub screw is located at the bottom of the rotor on the
marked side. To remove the rotor, turn counter- clockwise to loosen the
screw and then gently pull the rotor up vertically. Insert the new rotor
from the top onto the motor shaft and tighten the screw again. Turn
clockwise to tighten the screw which is located under the bottom of
rotor. Please check that the rotor is tight enough to avoid any accidents.
Note:
Tightening the screw will be in clockwise direction and the loosening will be in
counter - clockwise direction
�-- Sticker Mark
----
�GrubScrew
------------ .
To_;,ghte1---Key
-c:; ,;
__,
+--
To loosen
Motor Shaft