3900001 Rev. C
Page 18 of 24
User Manual
Issue
02
/
February 2015
/
Shasta User Manual
Shasta High Voltage Power Supply
Electrical Connections
J1 High Voltage Output & J3 Filament/Grid Connector (3 Pin)
The high voltage, filament and grid connections are made with custom cable assemblies, provided with
the power supply. These are available in a variety of lengths and configurations, depending on the
application requirement and type of X-ray tube being driven. The possible options are below:
Part Number
Description
9200008
Shasta Standard HV Cable, 1m
9200009
Shasta Standard HV Cable, 2m
9200010
Shasta Standard HV Cable, 3m
9200011
Shasta Apogee HV Cable, 1m
9200012
Shasta Apogee HV Cable, 2m
9200013
Shasta Apogee HV Cable, 3m
PIN
NAME
IN/OUT
FUNCTION
J3.1 FILAMENT
Output
Grounded filament positive output
J3.2 FILAMENT RETURN
Output
Grounded filament negative output
J3.3 GRID BIAS(OPTIONAL)
Output
Optional grid bias output
J2 DC Power Connector (2 Way)
Fitted to unit:
PTR/Phoenix STLZ950/2-G-5.08-H-Green
Customer part:
PTR/Phoenix AKZ950/2-5.08-Green (Supplied with unit)
PIN
NAME
IN/OUT
FUNCTION
1
+24V DC INPUT
Input
Main 24V DC
2
GND
Input
Main supply 0V
J4 Monitor/Interlock Connector (4 Way)
Fitted to unit:
PTR/Phoenix STLZ950/4-G-5.08-H-Green
Customer part:
PTR/Phoenix AKZ950/4-5.08-Green (Supplied with unit)
PIN
NAME
IN/OUT
FUNCTION
1
MONITOR RET/GND
Output
Return/ground for all monitor measurements
2
kV MONITOR
Output
High voltage kV monitor
3
mA MONITOR
Output
Beam current monitor
4
INTLK
Output
Interlock control connection