LEVEL-METER DAUGHTER BOARD
Original Instructions
6
Remove the screw that secures the blanking panel in the location where the board
is to be fitted. Keep the screw.
Figure 16-2
Daughter-board blanking-panel
7
Using appropriate ESD precautions, fit the level-meter board in the allocated
expansion slot.
8
Secure the board using the screw that was removed in step
.
9
Refit the board-clamp, using the two screws that were removed earlier.
10
Fit the lid back onto the unit, using the screws that were removed earlier.
16.2.2
Connecting the sensor
Connect the sensor to the appropriate 9-way D-connector on the rear panel of the
board. The pin connections are listed in
.
Table 16-1
Level-meter sensor pin connections
Pin
Signal Name
Helium Probe
Nitrogen Probe
1
VHIGH
V sense (Top)
n/c
2
VLOW
V sense (Bottom)
n/c
3
(Unused)
4
FREQ IN
Link to 5
OUTPUT FREQ (0 to 12 V)
5
FREQ OUT
Link to 4
n/c
6
I
HIGH
I (Top)
n/c
7
I
LOW
I (Bottom)
0 V
8
+12 V
n/c
+12V (20 mA maximum
9
CHASSIS GND
Screen
Screen