Oxford Instruments Andor Zyla Скачать руководство пользователя страница 2

MAN-06734

MAN-09086
ZQSG 0921 R1

Install the Camera Controller (PCIe) Card 

Troubleshoot

4

Connect the Camera and Switch it on

5

?

1. Unplug all cables from the rear of the computer.

2. Open the PC enclosure to gain access to the expansion slots.

3. Locate a suitable PCIe slot (x4, x8 x16 PCIe recommended - note that 

x1 slots may be shared with other devices resulting in reduced system 
performance). 

4.  If you are unsure which slot is correct, please consult the PC user manual.

5. 

Remove the filler bracket corresponding to the slot you intend to use.

6. Remove the controller card carefully from its protective ESD packaging and 

insert the card connector fully into the expansion slot.

7. 

Ensure the card’s mounting bracket is flush with any other mounting or filler 

brackets to either side of it, then secure the controller card in place.

8. Replace the computer cover and secure with mounting screws if applicable.

9. Reconnect any accessories you were using previously.

•  For

 Water Cooled 

models, ensure the coolant system is 

connected before the camera is switched on- 

please refer to the 

user manual.

•  Insert the 12 V DC power cable into the power connector at the 

rear of the camera, ensure the orientation is correct. 

NEVER

 

forcibly insert the connector.

•  Connect the supplied USB 3.0 cable between the USB 3.0 

connector on the camera and the corresponding slot on the 
Framegrabber PCIe card.

• 

If required, connect the I/O expander cable to the I/O 15 way D 

type connector.

•  Switch the camera ON using the ON/OFF switch. You should hear 

an audible confirmation (a long beep).

•  Turn ON the PC.
•  Start your software.
• 

View and adjust the acquisition settings as required.

• 

Start acquiring data.

Unused filler brackets

Gently press the
card into the slot

PCIe
Slot

Secure the card

to the PC

USB 3.0

Connection

12V     5A

Max

OFF

ON

I/O

USB

3.0

Made in UK

Zyla USB 3.0 version Connections

ON/OFF 
Switch

USB 3.0 
connector 
(locking)

I/O 15-

Way

D Type

Connector

12 V DC

Power 

Connector

Notes:
•  Use only the USB 3.0 cable supplied by Andor.
•  Do not connect any other device to the USB 3.0 PCIe card as this may interfere with normal camera 

operation or performance.

•  Camera operation with USB 3.0 PCIe cards not supplied by Andor cannot be guaranteed.

Note: If you need to adjust the C-mount, refer to the Zyla user manual.

Ensure your PC is powered off.

Ensure anti-static precautions and use the anti-static strap provided to prevent 
potential ESD damage. Keep all screws safe. 

Camera buzzer does not sound on start-up

•  The camera buzzer should be audible momentarily (as a short beep followed by a long beep) when the 

camera is switched to ON.

•  If this does not occur, ensure that power is connected to the camera and the ON/OFF switch is set to ON.

Camera is not recognized by PC

•  Ensure camera is switched to ON (The camera must be powered up before the software is initiated). 
•  Check that the camera cable(s) are securely and correctly connected between the camera head and the 

PCIe card in the PC (shown in step 4).

•  Check that the PCIe card and drivers have been installed properly 

(Control Panel – System – Hardware – Device Manager – Imaging Devices).

Please see the Hardware Guide for further information on Troubleshooting. 
Online access to technical documentation and software downloads is available at:  

 

andor.oxinst.com/downloads

Check & Set BIOS options (for PCs not supplied by Andor)

Enter the BIOS menu when starting PC. For Dell workstations, press F12 at start-up and select System Setup in the 
One Time Boot Menu.  For Dell workstations 3 options in the Performance menu of the BIOS need to be checked/
set:  
•  C-States Control – Disable C-States.
•  Intel Speed-step – Disable Speed-step. 
•  Memory Node Interleaving – Set from NUMA to SMP.  Note: This option is only available on larger workstations 

with 2 physical processors and may have a different name- ensure that NUMA is disabled.

Install the USB 3.0 PCIe Card Driver manually (not required for Windows 10)

USB 3.0 PCIe card drivers must be installed manually for PCs running versions prior to Windows 10.  
See Startech’s website: www.startech.com/downloads enter product code PEXUSB3S25 and hit Search. If a later 
driver is available, download and install.

Disable Power Management for the “Renesas USB 3.0 Host Controller”:

1. Navigate to ‘Renesas USB 3.0 Host Controller’ properties menu in 

Device Manager

 (Control Panel>Device 

Manager> Universal Serial Bus Controllers). Select the 

Advanced

 tab.

2. Check the 

Disable USB 3.0 power management functions 

checkbox.

Ensure that the Camera ON/OFF switch is in the OFF position AND your PC is switched OFF.

Отзывы: