4.Advanced User Guidebook
51
Chk Error
Even-Odd
:
Select Even or Odd.
Common Baud:
adjust
M
knob
to choose
common baud.
Custom Baud:
adjust
M
knob
to choose
baud
, ranges from 50 to 10,000,000.
Data
Data Bits
:
Set as 5
、
6
、
7
、
8 bits.
Data
:
Set data according to data bits,
ranges from 0-31, 0-63, 0-127 or 0-255.
Mode
Holdoff
Auto
Normal
Single
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one
waveform then stop
2. I2C Trigger
The I2C serial bus consists of SCL and SDA. The transmission rate is
determined by SCL, and the transmission data is determined by SDA. As
shown in below figure, oscilloscope can trigger on the start, restart, stop, ack
lost, specific device address or data value, also device address and data
value at the same time.
In I2C bus trigger mode, the trigger setting information is displayed on
bottom right of the screen, for example,
, indicates that
trigger type is I2C, CH1 trigger level is 0.00mV.
I2C Trigger menu
list:
MENU
SETTING
INSTRUCTION
Bus Type
I2C
Set vertical channel bus type as I2C trigger.
Source
SCL
SDA
Set SCL.
Set SDA.
When
Start
Trigger when SDA data transitions from
high to low while SCL is high.
Restart
When another start condition occurs before
a stop condition.