Figure 4-6 Bartlett window
Notes for using FFT
Use the default
dB
scale for details of multiple frequencies, even if they have very
different amplitudes. Use the
Vrms
scale to compare frequencies.
DC component or offset can cause incorrect magnitude values of FFT waveform. To
minimize the DC component, choose AC Coupling on the source signal.
To reduce random noise and aliased components in repetitive or single-shot events, set
the oscilloscope acquisition mode to average.
What is Nyquist frequency?
The Nyquist frequency is the highest frequency that any real-time digitizing oscilloscope
can acquire without aliasing. This frequency is half of the sample rate. Frequencies above
the Nyquist frequency will be under sampled, which causes aliasing. So pay more
attention to the relation between the frequency being sampled and measured.
Use Vertical Position and Scale Knobs
1.
The
Vertical Position
knob is used to adjust the vertical positions of the
waveforms.
The analytic resolution of this control knob changes with the vertical division.
37
Содержание XD3204AE
Страница 12: ......
Страница 13: ......
Страница 14: ......
Страница 15: ......
Страница 24: ......
Страница 25: ......
Страница 26: ......
Страница 27: ......
Страница 28: ......
Страница 29: ......
Страница 30: ......
Страница 31: ......
Страница 38: ......
Страница 42: ...Figure 4 2 Rectangle window Figure 4 3 Blackman window 35 ...
Страница 43: ...Figure 4 4 Hanning window Figure 4 5 Kaiser window 36 ...
Страница 46: ......
Страница 85: ......
Страница 99: ......
Страница 105: ......
Страница 107: ......
Страница 109: ......
Страница 110: ......
Страница 114: ......
Страница 115: ......
Страница 119: ......
Страница 120: ......
Страница 131: ......
Страница 132: ......
Страница 133: ......
Страница 134: ......
Страница 135: ......
Страница 136: ......
Страница 143: ......