To avoid possible human body injury, and / or device damage, and / or its accessory or communicating
facility damage, before working the device, Lilliput strongly recommend to read the following safety
information,
Warning
To avoid any potential short circuit or electric shock, DO use the power adapter of original local standard,
or recommended by Lilliput.
Warning
The channels of the device is non-isolated electrically. When working the device, to avoid possible short
circuit, the ground of two probes are NOT allowed to connect to 2 different non-isolated DC level.
The illustration of the device
’s built-in ground wire connection -
Ground Clip
Signal Input
Electrical Outlet
Probe
Device
AC Adapter
When the device communicating with PC via USB communication interface (with PC powered by AC
power source), the illustration of the ground wire connection -
Ground Clip
Signal Input
Device
(PC USB power)
PC
Electrical Outlet
Probe
USB Cable
It is NOT allowed to measure AC power when the device been powered by AC power source through the
adapter, or when the device powered by PC through USB connection cable (via USB communication
interface of PC, with PC powered by AC power source).
Содержание VDS6000 Series
Страница 1: ...VDS6000 Series 2CH PC Oscilloscopes User Manual www owon com ...
Страница 12: ...12 click Yes to continue and following window comes dot tick I accept the above license agreement then Next ...
Страница 14: ...14 to select additional items you may wish to install via scroll bar to select more necessary options ...
Страница 16: ...16 to welcome Review window via Next to get access to NI VISA installation one more Next ...